Cook G46218Recalls & reported problems
Cook Incorporated · Cook, Cutting Loop 24FR, .012" Wire · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002462182), received between Aug 7, 2017 and Apr 16, 2019 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break4
- Material Fragmentation2
- Fracture1
Patient problems
- No Consequences Or Impact To Patient4
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2019Break; Break
IT WAS INITIALLY REPORTED, DURING A TRANSURETHRAL RESECTION OF BLADDER TUMOR (TURBT) USING STORZ RESECTOSCOPE, THE COOK CUTTING LOOP BECAME LOOSE BUT NOT DETACHED DURING THE PROCEDURE. NO HARM TO PATIENT. OPENED ANOTHER CUTTING LOOP AND FINISHED PROCEDURE. THE DEVICE FAILURE ANALYSIS PERFORMED BY THE MANUFACTURER ON 01APR2019, FOUND THE RETURNED CUTTING LOOP WAS BROKEN, NOT LOOSE AS WAS INITIALLY REPORTED. AS REPORTED, NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED AND THERE WERE NO ADVERSE EFFECTS ON THE PATIENT AS A RESULT OF THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8520936 ↗MalfunctionNov 29, 2018Fracture
IT WAS REPORTED DURING THE TRANSURETHRAL RESECTION OF BLADDER TUMOR (TURPT) PROCEDURE , THE CUTTING LOOP WAS FRACTURED. THERE WERE NO ADVERSE EFFECTS REPORTED DUE TO THE EVENT AND NO REPORT OF AN ADDITIONAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8116156 ↗MalfunctionNov 28, 2018Break
IT WAS REPORTED THAT DURING THE RESECTION PORTION OF THE PROCEDURE, THE CUTTING LOOP BROKE IN THREE INSTANCES. THERE WERE NO ADVERSE EFFECTS REPORTED DUE TO THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8112444 ↗InjuryAug 8, 2017Break; Material Fragmentation; Break; Material Fragmentation
THE CUSTOMER REPORTED DURING A HYSTEROSCOPY PROCEDURE WHILE USING THE COOK CUTTING LOOP, THE RESECTOSCOPE LOOP BROKE OFF IN THE PATIENT. THE BROKEN PORTION WAS RETRIEVED FROM THE PATIENT WITHOUT CONSEQUENCE. A SECOND COOK CUTTING LOOP WAS USED. THE SECOND RESECTOSCOPE LOOP BROKE OFF BUT IT COULD NOT BE FOUND. AS REPORTED, THESE ISSUES CONTRIBUTED TO AN EXTENDED TIME IN SURGERY. THE LENGTH OF TIME THE SURGERY WAS EXTENDED HAS NOT BEEN PROVIDED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THESE ISSUES. THIS REPORT IS BEING FILED TO CAPTURE THE REPORTED FIRST LOOP BREAKAGE. THE SECOND LOOP BREAKAGE CAN BE FOUND IN MEDWATCH REPORT NUMBER 1820334-2017-02296.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6773276 ↗InjuryAug 7, 2017Break; Material Fragmentation; Break; Material Fragmentation
THE CUSTOMER REPORTED DURING A HYSTEROSCOPY PROCEDURE WHILE USING THE COOK CUTTING LOOP, THE RESECTOSCOPE LOOP BROKE OFF IN THE PATIENT. THE BROKEN PORTION WAS RETRIEVED FROM THE PATIENT WITHOUT CONSEQUENCE. A SECOND COOK CUTTING LOOP WAS USED. THE SECOND RESECTOSCOPE LOOP BROKE OFF BUT IT COULD NOT BE FOUND. AS REPORTED, THESE ISSUES CONTRIBUTED TO AN EXTENDED TIME IN SURGERY. THE LENGTH OF TIME THE SURGERY WAS EXTENDED HAS NOT BEEN PROVIDED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THESE ISSUES. THIS REPORT IS BEING FILED TO CAPTURE THE SECOND LOOP BREAKAGE ISSUE. THE FIRST LOOP BREAKAGE REPORT CAN BE FOUND IN MEDWATCH REPORT NUMBER 1820334-2017-02159.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 6772678 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page