OEC ELITERecalls & reported problems
Oec Medical Systems, Inc. · MEDICAL DEVICE, UDI LABEL PLACEMENT, OEC ELITE · Class 2
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682121729), received between Mar 25, 2026 and Mar 25, 2026 — classified as 1 death.
4 FDA recall records name this device's identifier, 4 still open.
2 other recalls cover devices cleared under the same FDA submission (K170752, K171565, K172550, K192819); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 31, 2025Open, Classified
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Z-1239-2025 · Root cause: Software design · FDA record ↗
- Nov 21, 2024Open, Classified
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Z-0703-2025 · Root cause: Process control · FDA record ↗
- Nov 21, 2024Open, Classified
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Z-0704-2025 · Root cause: Process control · FDA record ↗
- May 15, 2023Open, Classified
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Z-2278-2023 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathMar 25, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, AN INTERNAL EMAIL ORIGINATING FROM A CUSTOMER SITE WAS SENT TO A GE HEALTHCARE (GEHC) SALES REPRESENTATIVE INDICATING THAT A PATIENT EXPIRATION HAD OCCURRED AND THAT THE EVENT WAS POTENTIALLY ASSOCIATED WITH A GE HEALTHCARE C ARM. THE CONTENT OF THE CUSTOMER¿S EMAIL DESCRIBED THE FOLLOWING SEQUENCE OF EVENTS: 1. THE PATIENT PRESENTED FOR A ROUTINE UROLOGIC PROCEDURE REQUIRING INTRAOPERATIVE FLUOROSCOPY. 2. THE GEHC C ARM EQUIPMENT WAS MANEUVERED TO OBTAIN THE DESIRED VIEW. 3. THE CLINICAL TEAM DID NOT RECOGNIZE AN IMAGE INVERSION AND SUBSEQUENTLY REQUESTED ADMINISTRATION OF CONTRAST DYE. 4. AN INCORRECT DOSE OF CONTRAST DYE WAS ADMINISTERED, WHICH THE CUSTOMER REPORTED AS LEADING TO ¿CATASTROPHIC OUTCOMES,¿ INCLUDING A REPORTED PATIENT EXPIRATION. NO ADDITIONAL INFORMATION REGARDING THE DATE OF THE PROCEDURE, SPECIFIC DEVICE SETTINGS, IMAGES, WORKFLOW STEPS, OR ANY DIRECT OBSERVATIONS FROM GEHC PERSONNEL HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 24687627 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170752, K171565, K172550, K192819. One submission often covers many models, so check the product and lot details in each record.
- Z-1819-2025 · Mar 21, 2025 · Open, ClassifiedOEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during d
- Z-2161-2021 · Jun 10, 2021 · TerminatedOEC Elite mobile C-arm - Product Usage: intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during di
Other System, x-ray, fluoroscopic, image-intensified devices
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified9 recalls7 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified7 recalls1 report
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified6 recalls4 reports
- ProxiDiagnost N90 706110Philips Medical Systems DMC GmbH · System, X-Ray, Fluoroscopic, Image-Intensified5 recalls1 report
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- CombiDiagnost R90 709031Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls11 reports
- ProxiDiagnost N90 706100Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls6 reports
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000CCanon Inc. · System, x-ray, fluoroscopic, image-intensified4 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.