Skip to content

CombiDiagnost R90 709031Recalls & reported problems

Philips Medical Systems DMC GmbH · The CombiDiagnost R90 is a multi-functional general Radiography/Fluoroscopy (R/F) system.that includes - Adjustable Straps for Patient Support; - Adult headrest; - Ankle Clamps; - Arm Support; - Babix holder vertical stand; - Compression Belt (singles. pull); - Console trolles; - Detector Holder for Patient Bed; - Extra Stitching Ruler; - Hand grips (Pair of); - Handle; - Infusion bottle holder; - Leg Supports (Pair of); - Mattress GXR Systems; - Mobile Detector Holder; - Monitor Ceiling Suspension(Family); - Monitor CS for1 Monitor; - Monitor CS for2 Monitors; - Monitor CS for3 Monitors; - Monitor Trolley Family: - Monitor Trolley for 1 Monitor assembly; - Monitor Trolley for 2 Monitors assembly; - Patient Support for Stitching; - Rotable stool for footrest; - Shoulder supports (Pair of); - Side Bar; - Stretch Grip for VS/VM Stand · Class 2

All models of CombiDiagnost R90 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
4

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00884838101456), received between Sep 29, 2021 and Sep 29, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 1, 2025Class IIOpen, Classified

    Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

    Z-0462-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Sep 21, 2023Class IIOpen, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0271-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Feb 11, 2022Class IIOpen, Classified

    While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

    Z-1088-2024 · Root cause: Software design · Recall details · FDA record ↗

  4. Oct 21, 2021Class IIOpen, Classified

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

    Z-1355-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Patient Data Problem2
  • Display or Visual Feedback Problem1
  • Insufficient Device Problem Information1
  • Mechanical Jam1
  • Moisture Damage1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Unspecified Musculoskeletal problem1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Detachment of Device or Device Component

    It was reported that the detachable foot table board was broken and would not attach to the table due to physical damage. The device was not in clinical use, and there was no patient or user harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934069 ↗
  • InjuryDec 15, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE CEILING SUSPENSION (CS) WAS DIFFICULT TO MOVE, AND AS A RESULT, THE USER EXPERIENCED A BACK INJURY. THE DEVICE WAS IN CLINICAL USE.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 23800305 ↗
  • MalfunctionOct 17, 2025Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE CRACK IN THE MOUNTING POSITION OF THE ACTUATOR. THE USE OF DEVICE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23321489 ↗
  • MalfunctionApr 23, 2025Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THAT ON COMBIDIAGNOST R90 INDICATING THAT TABLE UP DOWN MOVEMENT IS NOT STRAIGHT INTERMITTENTLY AND MIGHT CAUSE HARM TO THE PATIENT. THERE WAS NO ACTUAL HARM REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21891546 ↗
  • MalfunctionMar 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE FOOTSTEP WHERE THE PATIENT CAN STAND ON IT HAS COME OFF / SEEMS TO BE BROKEN. THE DEVICE WAS IN CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21733383 ↗
  • MalfunctionNov 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DURING PREPARATION OF EXAMINATION, WHEN THE TABLE WAS BEING TITLED DOWNWARDS THE FOOT STAND FELL AND BROKE. THERE WAS NO USER IMPACT AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20804335 ↗
  • MalfunctionNov 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE WAS IN CLINICAL USE WHEN THE FOOTSTEP OF THE TABLE WHERE THE PATIENT STANDS HAS COME AND SEEMS TO BE BROKEN. THERE WAS NO PATIENT OR USER IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20607808 ↗
  • MalfunctionAug 9, 2023Moisture Damage; Moisture Damage

    THE CUSTOMER REPORTED THAT WATER WENT INTO CABLES OF SYSTEM AFTER A WATER PIPE FROM THE BATHROOM BURST. THE CUSTOMER REQUESTED A CHECK OF THE SYSTEM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17487194 ↗
  • MalfunctionDec 28, 2021Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT ONE IMAGE OF A PATIENT WAS DISPLAYED IN AN OTHER PATIENT'S FOLDER. THERE WAS NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13099197 ↗
  • MalfunctionDec 6, 2021Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT ONE IMAGE OF A PATIENT WAS DISPLAYED IN AN OTHER PATIENT'S FOLDER. THERE WAS NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12934779 ↗
  • MalfunctionSep 29, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THAT THE CURRENT IMAGE IS MIXED WITH THE PREVIOUS IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12547062 ↗

Other System, x-ray, fluoroscopic, image-intensified devices

All System, x-ray, fluoroscopic, image-intensified devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page