CoolSeal™ Trinity CSL-TR105-30Recalls & reported problems
Hologic, Inc. · Maryland Laparoscopic Sealer/Divider/Dissector, 30 cm · Class 2
All models of CoolSeal Trinity (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00850346007030), received between Feb 14, 2022 and Aug 1, 2025 — classified as 1 death, 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Material Fragmentation1
- Peeled/Delaminated1
Patient problems
- Hemorrhage/Blood Loss/Bleeding3
- Foreign Body In Patient1
- Intra-Abdominal Hemorrhage1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 1, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT DURING A COOLSEAL PROCEDURE ON (B)(6) 2025, DURING THE MIDDLE OF THE PROCEDURE WHILE PERFORMING A LAPAROSCOPIC DIAPHRAGMATIC HERNIA ON A 1 YEAR OLD, THE STAFF NOTICED A BLACK MATERIAL ON THE ABDOMEN COMING FROM THE DEVICE. THE COOLSEAL WAS CHECKED AND IT WAS NOTICED THAT THE COATING CLOSE TO THE TIP WAS PEELING. THE DEVICE WAS EXCHANGED FOR A NEW DEVICE AND THE PROCEDURE COMPLETED SUCCESSFULLY. MATERIAL WAS OBSERVED INSIDE THE CAVITY AND IT WAS IRRIGATED AND FLUSHED. NO ADDITIONAL MEDICAL INTERVENTION WAS REPORTED. NO OTHER INFORMATION AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 22692235 ↗DeathApr 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A COOLSEAL TRINITY PROCEDURE ON (B)(6) 2024, A COOLSEAL TRINITY 30CM AND COOLSEAL GENERATOR WERE USED DURING A THORACOSCOPIC PROCEDURE ON A CHILD (6 YEARS OLD, FEMALE), DISSECTING AND SEALING VESSELS. DURING THE PROCEDURE IT WAS IDENTIFIED THAT THE PATIENT WAS BLEEDING FROM A PREVIOUSLY DISSECTED AND SEALED BRANCH OF THE PULMONARY ARTERY, AND THERE WERE CONCERNS OF A POTENTIAL SEAL FAILURE. THE PATIENT SUFFERED SIGNIFICANT BLEEDING WHICH COULD NOT BE STOPPED, RESULTING IN PATIENT DEATH. ON A FOLLOW UP EMAIL IT WAS DETERMINED THAT THE BLEEDING STARTED 3 HOURS AFTER THE SURGERY BEGAN, MAJOR BLEEDING WAS ENCOUNTERED FROM A CORNER OF A PREVIOUSLY SEALED AND DIVIDED BASAL ARTERIAL BRANCH. THE PROCEDURE WAS A VIDEO ASSISTED THORACOSCOPIC LOBECTOMY SURGERY. NO PRE-EXISTENT CONDITIONS WERE REPORTED. MULTIPLE STEPS WERE TAKEN AFTER THE BLEEDING WAS OBSERVED TO TREAT IT, COMPRESSION, TRANEXAMIC ACID AND BLOOD TRANSFUSION. NO OTHER INFORMATION WAS RECEIVED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21886742 ↗InjuryJan 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT RECEIVED A COOLSEAL PROCEDURE ON (B)(6) 2024 ON A SALPINGECTOMY. THE PATIENT WAS REPORTED TO HAVE AN ACTIVE ABDOMINAL BLEEDING AFTER THE PROCEDURE WHICH REQUIRED TO BE TAKEN TO THE OPERATING ROOM. CAUSE OF BLEEDING IS NOT KNOWN. PATIENT WAS LATER RELEASED FROM THE HOSPITAL. NO OTHER INFORMATION AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21193017 ↗InjuryJul 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2024 A RIGHT LOWER LUNG RESECTION/LOBECTOMY WITH A COOLSEAL TRINITY PROCEDURE WAS PERFORMED AND THE SURGEON WENT TO SEAL THE PULMONARY ARTERY, LOCKED JAWS, AND SEALED USING THE TRIPLE SEAL TECHNIQUE BEFORE CUTTING. THE DOCTOR HAD COMPLETE SEAL TONES BUT SHORTLY AFTER THE PULMONARY ARTERY BLED AND THE SURGEON COULD NOT GET THE BLEEDING UNDER CONTROL. THE DOCTOR HAD TO IMMEDIATELY OPEN THE PATIENT UP AND FINISH THE CASE WITH OTHER MODALITIES. THE PATIENT DIDN'T REQUIRE ANY ADDITIONAL MEDICATION. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19837713 ↗MalfunctionOct 20, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A COOLSEAL PROCEDURE ON (B)(6) 2023, THE PHYSICIAN WAS PERFORMING AN OVARIAN SURGERY AND IN THE MIDDLE OF THE PROCEDURE IT WAS NOTICED THAT A PIECE OF THE OF THE COOLSEAL BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY RETRIEVE THE PIECE OF THE DEVICE AND NO PATIENT INJURY WAS REPORTED. PROCEDURE WAS COMPLETED WITH SAME DEVICE AND NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17981243 ↗InjuryFeb 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER ENCOUNTERED SIGNIFICANT BLEEDING WHILE USING COOLSEAL TRINITY.
Patient problems: Intra-Abdominal Hemorrhage
MAUDE report 13524232 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page