Skip to content

Lithotripsy Probe, Sterile Single Use ShockPulse-SE, 10.2 Fr (3.40mm) X 396mm SPL-PD340Recalls & reported problems

Cybersonics Inc · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00855279005122), received between Sep 26, 2017 and Sep 26, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Metal Shedding Debris” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Metal Shedding Debris1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2017Metal Shedding Debris; Metal Shedding Debris

    DESCRIPTION PROVIDED BY DISTRIBUTOR. "SPL-PR340 PROBE (LOT NO. W1504338) IS USED IN A PCNL CASE PERFORMED BY UROLOGIST. THIS PHENOMENON OCCURRED DURING THE 2ND TIME OF USAGE. THE PROBE WAS UNPACKED AND ACTIVATED WITH HAND SWITCH BY A DOCTOR. DURING PROCEDURE, A SMALL GLITTER FRAGMENTS; PROBABLY A COATING OF SHOCKPULSE PROBES WAS PEEL OFF ON THE PATIENT'S KIDNEY. DOCTOR HAS REMOVED THOSE FRAGMENTS AND MOST OF THE FRAGMENTS HAVE BEEN REMOVED COMPLETELY EXCEPT FOR VERY SMALL PIECES. THE RESULT OF THE OPERATION / SURGERY WAS COMPLETED SAFELY. AFTER THE CASE, THE SUBJECTED PRODUCT HAS BEEN FOUND THAT THE SURFACE OF PROBE WAS SCRATCHED ALTHOUGH THIS CASE WAS JUST THE SECOND TIME OF USING". MODEL NO. OF PROBES MENTIONED IN COMPLAINT FROM DISTRIBUTOR: SPL-PR340 (REUSABLE). ACTUAL MODEL NO. OF PROBES SHIPPED TO DISTRIBUTOR: SPL-PDBX340 (SINGLE USE, DISPOSABLE PROBES).

    Patient problems: Foreign Body In Patient

    MAUDE report 6893665 ↗

Other Lithotriptor, Ultrasonic devices

All Lithotriptor, Ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page