Tricera™ Handpiece, Autoclavable R-10023Recalls & reported problems
Relign Corporation · Shaver Handpiece, Autoclavable · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00859273007236), received between May 11, 2022 and Mar 10, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device2
- Patient Device Interaction Problem2
- Sparking2
- Complete Loss of Power1
- Difficult to Remove1
- Electrical /Electronic Property Problem1
- Failure to Cut1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Burn(s)2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2025Sparking; Sparking
IT WAS REPORTED THAT DURING SURGERY THE DEVICE THREW SPARKS WHEN COMING INTO CONTACT WITH OTHER METAL AND THEN IT STOPPED WORKING. THERE WAS NO PATIENT IMPACT OR DELAY. DUE DILIGENCE IS COMPLETE AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21559377 ↗MalfunctionSep 19, 2024Difficult to Remove; Failure to Cut; Difficult to Remove; Failure to Cut
IT WAS REPORTED THAT BEFORE SURGERY WHEN THE DEVICE WAS PLUGGED IN TO TEST IT PRIOR TO PATIENT USE, THE CAUTERY WAS NOT WORKING, NOT CUTTING. THE STERILIZATION CAP CAN 'T BE REMOVED WITHOUT PUTTING IT IN A VICE. THERE WAS NO PATIENT INVOLVEMENT. DURING DEVICE EVALUATION, IT WAS DISCOVERED THAT THERE WERE NO METAL WIRES EXPOSED OUTSIDE OF THE DEVICE UPON ARRIVAL, THE EXPOSED WIRE WAS FOUND AFTER DISASSEMBLY TO FURTHER INVESTIGATE THE FAILURE. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20260146 ↗InjuryOct 30, 2023Overheating of Device; Patient Device Interaction Problem; Overheating of Device; Patient Device Interaction Problem
IT WAS REPORTED THAT DURING SURGERY, THE 3 IN 1 SHAVER OVERHEATED AND BURNED THE PATIENTS ARM. THE PATIENT HAD A BURN IN THE SHAPE OF THE SHAVER ON THEIR ARM. THERE WAS NO DELAY IN THE PROCEDURE. DUE DILIGENCE IS IN PROCESS AND THERE IS NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Burn(s)
MAUDE report 18030403 ↗InjuryOct 13, 2023Overheating of Device; Patient Device Interaction Problem; Overheating of Device; Patient Device Interaction Problem
THERE IS NO ADDITIONAL INFORMATION AVAILABLE REGARDING THE EVENT.
Patient problems: Burn(s)
MAUDE report 17931466 ↗MalfunctionSep 8, 2022Electrical /Electronic Property Problem; Mechanical Problem; Smoking; Suction Problem; Sparking; Electrical /Electronic Property Problem; Mechanical Problem; Smoking; Suction Problem; Sparking
IT WAS REPORTED THE DEVICE WASN'T WORKING CORRECTLY AND THE SUCTION WAS OFF, SO THEY PULLED IT OUT OF THE PATIENT AND HIT THE SHAVER MODE TO CHECK IT OUT. THE DEVICE SPARKED WHILE OSCILLATING. THE HANDPIECE WAS SWITCHED OUT AND EVERYTHING WORKED GREAT AFTER THAT AND PATIENT WAS NOT HARMED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15385948 ↗MalfunctionMay 11, 2022Complete Loss of Power; Complete Loss of Power
IT WAS REPORTED THAT THE RF FUNCTION STOPPED WORKING IN THIS HANDPIECE DURING THE MIDDLE OF THE SURGERY AND THE DOCTOR WAS NOT ABLE TO USE ABLATION OR COAGULATION. NO HARM OR DELAY REPORTED. EVALUATION OF THE DEVICE LATER FOUND THAT THE PCBA AND SURROUNDING RTV WAS FOUND TO BE BURNT AND THE SURFACE ON THE ELECTRODE RING ALSO SHOWED EVIDENCE OF BURNING. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14357068 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.