CombiDiagnost R90 709030Recalls & reported problems

Philips Medical Systems DMC GmbH · Image-intensified fluoroscopic x-ray system · Class 2

All models of CombiDiagnost R90 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884838076747), received between Aug 12, 2019 and Aug 26, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Device Fell” (5 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 21, 2023Class IIOpen, Classified

    Potential for units suspended on the ceiling to fail and fall.

    Z-0271-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 11, 2022Class IIOpen, Classified

    While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

    Z-1087-2024 · Root cause: Software design · Recall details · FDA record ↗

  3. Oct 21, 2021Class IIOpen, Classified

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.

    Z-1355-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Fell5
  • Device Handling Problem4
  • Detachment of Device or Device Component2
  • Device Emits Odor1
  • Manufacturing, Packaging or Shipping Problem1
  • Poor Quality Image1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2025Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    IT WAS REPORTED THAT THE REMOTE ALERT OF A MECHANICAL HEAVY SHOCK TO THE DETECTOR. THE DEVICE WAS NOT IN CLINICAL USE AND THERE WAS NO HARM TO PATIENT OR USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892437 ↗
  • MalfunctionMay 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BASE WHERE THE PATIENT WAS PLACED HAD BROKEN. THE DEVICE WAS IN CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM. ADDITIONAL INFORMATION RECEIVED THAT THE WHEEL INSIDE THE FOOTREST WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069033 ↗
  • MalfunctionOct 12, 2024Detachment of Device or Device Component; Device Fell; Detachment of Device or Device Component; Device Fell

    IT WAS REPORTED THAT DURING CLINICAL USE, WHEN SETTING UP THE ROOM FOR EXAMINATION FOR A THORAX EXAMINATION WHERE THE TABLE WAS VERTICAL WITH SOURCE IMAGE DISTANCE OF 150 CM, THE STAND GEOMETRY OF THE SYSTEM FELL DOWN ON THE FLOOR. THE CHAIN IN THE STAND GEOMETRY SNAPPED WHEN THE USER WAS ATTEMPTING TO TILT UPWARDS AT 5 DEGREE ANGLE RESULTING IN FALLING DOWN OF THE COLLIMATOR AND THE TUBE. IT WAS CONFIRMED THAT THERE WAS NO IMPACT OR INJURY TO THE PATIENT AND THE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20435005 ↗
  • MalfunctionApr 26, 2024Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    THE CUSTOMER REPORTED THAT THE PORTABLE DETECTOR FAILED THE GAIN CALIBRATION DURING THE QUARTERLY CHECK. A DROPPED DETECTOR CAN LEAD IN WORST CASE TO A FRACTURE IN THE FOOT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19191028 ↗
  • MalfunctionAug 23, 2023Device Handling Problem; Device Fell; Device Handling Problem; Device Fell

    THE CUSTOMER REPORTED THAT THE PORTABLE DETECTOR HAS FALLEN DOWN. A DROPPED DETECTOR CAN LEAD IN WORST CASE TO A FRACTURE IN THE FOOT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17603346 ↗
  • MalfunctionAug 8, 2023Poor Quality Image; Device Handling Problem; Device Fell; Poor Quality Image; Device Handling Problem; Device Fell

    THE CUSTOMER REPORTED THAT THE PORTABLE DETECTOR FELL AND HAS NOW ARTEFACTS IN THE IMAGE. A DROPPED DETECTOR CAN LEAD IN WORST CASE TO A FRACTURE IN THE FOOT. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17480266 ↗
  • MalfunctionMar 15, 2023Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS IDENTIFIED THAT AN OBSOLETE SPDU (SYSTEM POWER DISTRIBUTION UNIT) SPARE PART IS INSTALLED IN A COMBIDIAGNOST SYSTEM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16544522 ↗
  • MalfunctionOct 8, 2019Device Emits Odor

    THE CUSTOMER COMPLAINED ABOUT A BURNING SMELL FROM THE SYSTEM. NO INJURY OCCURRED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9166537 ↗
  • MalfunctionAug 12, 2019Smoking

    THE CUSTOMER COMPLAINED ABOUT THE SYSTEM POWER DISTRIBUTION UNIT (IN SHORT NAMED SPDU) PRODUCED A SIGNIFICANT AMOUNT OF SMOKE. HOSPITALS SMOKE SENSOR WAS ACTIVATED AND THE FIRE DEPARTMENT WAS INVOLVED.. NO PATIENT INVOLVEMENT

    Patient problems: No Patient Involvement

    MAUDE report 8883839 ↗

Other System, x-ray, fluoroscopic, image-intensified devices