Xper Flex Cardio Physiomonitoring System Xper Flex Cardio 2020 Rev D ExchangeRecalls & reported problems

Philips Ds North America LLC · Xper Flex Cardio 2020 Rev D Exchange · Class 2

All models of Xper Flex Cardio Physiomonitoring System (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838086012), received between Apr 7, 2023 and Apr 7, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Read Input Signal” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Read Input Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2023Failure to Read Input Signal; Failure to Read Input Signal

    THE CUSTOMER REPORTED THAT THE FC2010 STOPPED POPULATING VITALS DURING CASE WHILE PATIENT WAS CODING. THE CUSTOMER REPORTED ECG TRACING WAS NOT AVAILABLE, NO NON-INVASIVE MEASUREMENTS (BP, HR, SPO2), AND NO ARTERIAL TRACINGS WERE FUNCTIONAL AND DISPLAYED. THE FC2010 WAS NOT CONNECTED TO OTHER DEVICES. THE PATIENT CONTINUED TO BE APPROPRIATELY RESUSCITATED. THE IMPELLA (VENTRICULAR ASSIST DEVICE) WAS USED FOR THE PATIENT'S ARTERIAL TRACING. THIS DID NOT LEAD TO HARM TO THE PATIENT AS THE STAFF WAS ABLE TO SEE ARTERIAL TRACINGS USING THE IMPELLA DEVICE WHILE TROUBLESHOOTING THE FLEX CARDIO FOR THREE MINUTES. APPROPRIATE RESUSCITATION CONTINUED, BUT THE USER WAS MORE CONCERNED WITH PROPER DOCUMENTATION OF THE VITAL SIGNS WITH REGARD TO THE DEVICE ISSUE. THERE WAS NO INDICATION THIS ISSUE RESULTED IN ANY ADDITIONAL HARM TO THE PATIENT. A PHILIPS REMOTE SERVICE ENGINEER (RSE) INTERVIEWED THE CUSTOMER TO EVALUATE THE DEVICE IN QUESTION. THE RSE PROVIDED THE CUSTOMER WITH THE FC2010 PREVENTIVE XPER IFU DOCUMENTATION TO CUSTOMER TO REVIEW. THE RSE RECOMMENDED, IF POSSIBLE, TO POWER DRAIN THE FC2010 TO RESTART THE FC2010\HEMO APP STATION DAILY, OR MONDAY MORNINGS, BEFORE EACH CASE. THE RSE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699847 ↗

Other Monitor, physiological, patient (without arrhythmia detection or alarms) devices

All Monitor, physiological, patient (without arrhythmia detection or alarms) devices →