Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838102804), received between Dec 9, 2022 and Dec 9, 2022 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2022No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT A XPER FLEX CARDIO DEVICE WAS HAVING WAVE TRACINGS FREEZE AND INTERMITTENTLY SIGNALS. THE PHILIPS TECHNICAL CONSULTANT (TC) WAS UNABLE TO CONFIRM THE EVENT. THE TC WENT ON SITE AND RAN TESTING, BUT WAS UNABLE TO DUPLICATE THE EVENT. THE TC DID NOTE THAT THE SETTINGS NEEDED TO BE ADJUSTED FOR OPTIMAL PERFORMANCE. THE CAUSE OF THE EVENT WAS UNABLE TO BE DETERMINED. THE TC CORRECTED SETTINGS WHILE ON SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15947820 ↗

Other Monitor, physiological, patient (without arrhythmia detection or alarms) devices

All Monitor, physiological, patient (without arrhythmia detection or alarms) devices →