Xper Flex Cardio Physiomonitoring System FC2010 (REF 453564241901)Recalls & reported problems
Philips Ds North America LLC · Class 2
All models of Xper Flex Cardio Physiomonitoring System (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00884838103610), received between Feb 9, 2021 and Jul 5, 2022 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Computer Software Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Computer Software Problem2
- Application Program Freezes or Fails to Launch1
- Application Program Problem1
- Delayed Program or Algorithm Execution1
- Device Displays Incorrect Message1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 5, 2022Output Problem; Insufficient Device Problem Information; Output Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT DURING A DEFIB-IMPLANTATION PROCEDURE, THE ECG CURVES FLAT LINED AND THERE WAS NO SPO2 DISPLAYED. PATIENT OUTCOME IS CURRENTLY UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 14917225 ↗MalfunctionMar 15, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT SOMETIMES THE SWEEP SPEED CHANGES AUTOMATICALLY. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13770651 ↗MalfunctionJan 28, 2022Application Program Problem; Application Program Problem
THE CUSTOMER REPORTED THAT THE NON-INVASIVE BLOOD PRESSURE READINGS ARE DUPLICATING AND NOT TAKING NEW BLOOD PRESSURES EVERY FIVE MINUTES AS IT IS SET TO. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13385456 ↗MalfunctionJun 7, 2021Computer Software Problem; Computer Software Problem
THE CUSTOMER REPORTED THAT THE DEVICE ONLY SHOWS DUPLICATED NIBP (NON-INVASIVE BLOOD PRESSURE) MEASUREMENTS FOR EVERY CASE. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11952530 ↗MalfunctionMay 21, 2021No Audible Alarm; No Audible Alarm
THE CUSTOMER REPORTED SPO2 SPECIFICATIONS IN THE EXAMINATION ROOM WITHOUT SOUND. NO VITAL SIGNS SOUND FOR SPO2 MAKES IT DIFFICULT TO CONTINUE WORKING. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11866515 ↗MalfunctionMay 6, 2021Computer Software Problem; Computer Software Problem
THE CUSTOMER REPORTED THE SWEEP SPEED CHANGING AUTOMATICALLY DURING PROCEDURE. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11782593 ↗MalfunctionMay 4, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
THE CUSTOMER REPORTED AN ECG HANG ISSUE. THE DISPLAY IS FREEZING. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11769966 ↗MalfunctionApr 14, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REQUESTED TO HAVE THE DEVICE SWAPPED OUT DUE TO THE QRS COMPLEX BEING ¿CLIPPED¿ AT 187MMHG WHEN DOING FFR (FRACTIONAL FLOW RESERVE). THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11667474 ↗MalfunctionApr 14, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REQUESTED TO HAVE THE DEVICE UPGRADED DUE TO THE QRS COMPLEX BEING ¿CLIPPED¿ AT 187MMHG WHEN DOING FFR (FRACTIONAL FLOW RESERVE). THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT, BUT NO ADVERSE EVENT TO PATIENT OR USER WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11667280 ↗MalfunctionMar 4, 2021Delayed Program or Algorithm Execution; Delayed Program or Algorithm Execution
THE CUSTOMER REPORTED THAT THE SWEEP SPEED OF THE PROGRAM HAS BEEN CHANGING AUTOMATICALLY AND IT APPEARS THAT THE PATIENTS HR IS NOT BEING DISPLAYED PROPERLY. IT HAPPENED TWICE THIS MORNING. THE SYSTEM HAS TWO SCREENS ONE IN THEATRE AND ONE ON THE VIEWING ROOM. THE SCREEN IN THE VIEWING ROOM IS THE ONE WHICH HAS THE CHANGE IN THE SWEEP SPEED. THERE WAS NO REPORTED PATIENT IMPACT / INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11417816 ↗MalfunctionFeb 9, 2021Device Displays Incorrect Message
THE CUSTOMER REPORTED THAT THE HEART RATE VALUE WAS STUCK AT 99 AND NOT UPDATING. THE DEVICE WAS IN USE ON A PATIENT. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11301616 ↗
Other Monitor, physiological, patient (without arrhythmia detection or alarms) devices
- Rad-97 9738Masimo Corporation · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)1 death222 reports
- Xper Flex Cardio Physiomonitoring System Xper Flex Cardio FC2010 REV CPhilips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls1 death42 reports
- Xper Flex Cardio Physiomonitoring System FC 2010Philips Ds North America LLC · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 deaths38 reports
- Welch Allyn, Inc. 901058Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls5 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls3 reports
- Welch Allyn, Inc. 901060Welch Allyn, Inc. · MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS)2 recalls2 reports
- McKesson Cardiology Hemo 14.0Change Healthcare Canada Company · Monitor,Physiological,Patient(Without Arrhythmia Detection Or Alarms)2 recalls2 deaths2 reports
- Change Healthcare Cardiology Hemo 14.2Change Healthcare Canada Company · Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)2 recalls1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page