COBLATION 72290105SRecalls & reported problems
Smith & Nephew, Inc. · WEREWOLF GENERATOR SERVICE REPLACEMENT · Class 2
All models of Coblation (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00885556645352), received between Apr 22, 2019 and Feb 8, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Power Up3
- Communication or Transmission Problem2
- Electrical Shorting2
- Connection Problem1
- Device Alarm System1
- Device Emits Odor1
- Display or Visual Feedback Problem1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement5
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 8, 2026Device Alarm System; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, AFTER THE WEREWOLF CONTROLLER WAS TURNED ON, THE SCREEN GOT DARK, AND AN ALARM SOUNDED. THE SURGERY WAS INTERRUPTED AND WAS SCHEDULED FOR ANOTHER DAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24288347 ↗MalfunctionMar 25, 2024Electrical Shorting; Electrical Shorting
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION THE NURSE DID NOT CLAMP THE WATER BEFORE CHANGING THE BAGS OF FLUID HANGING ABOVE THE WEREWOLF CONTROLLER. WATER WAS POURED DIRECTLY INTO THE BACK OF THE DEVICE CAUSING THE BOX TO SPARK OUT THE BACK AND COMPLETELY FRY. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18969690 ↗MalfunctionFeb 11, 2021Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT THE GENERATOR WILL NOT TURN ON. NO CASE INVOLVED. THEREFORE, THERE WAS NO PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11319753 ↗InjuryDec 16, 2020Material Separation; Material Separation
IT WAS REPORTED THAT WEREWOLF CONTROLLER'S POWER BUTTON CAME APART FROM CONSOLE AND PUSHED INSIDE THE MACHINE, DURING SET UP OR INSPECTION FOR AN UNKNOWN PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR'S DEVICE. NO PATIENT INJURY OR DELAYS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11018756 ↗MalfunctionDec 9, 2020Electrical Shorting; Electrical Shorting
IT WAS REPORTED THAT THE, BEFORE A SHOULDER PROCEDURE, THE WEREWOLF CONTROLLER HAD A SMOKING SMELL COMING FROM IT. NO PATIENT INVOLVEMENT REPORTED. RESULTS OF INVESTIGATION REVEALED THE UNIT THE DOES NOT TURN ON WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10976938 ↗MalfunctionNov 22, 2020Connection Problem; Connection Problem
IT WAS REPORTED THAT THE WEREWOLF CONTROLLER HAD AN ERROR AFTER IT WAS REBOOTED AND THE HANDPIECE WAS NOT DETECTED. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10881673 ↗MalfunctionNov 13, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE WEREWOLF CONTROLLER COULD NOT READ THE HAND PIECE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10835377 ↗MalfunctionOct 15, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE WEREWOLF CONTROLLER WAS NOT RECOGNIZING THE WANDS. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10681986 ↗MalfunctionSep 9, 2020Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT, THE "WEREWOLF CONTROLLER" WOULD NOT TURN ON. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10511303 ↗MalfunctionJul 29, 2020Failure to Power Up; Failure to Power Up
IT WAS REPORTED THAT THE CONTROLLER DOES NOT TURN ON. NO CASE INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 10341606 ↗MalfunctionApr 22, 2019Device Emits Odor; Sparking
IT WAS REPORTED THAT THE DEVICE SMELLS LIKE IT IS BURNING AND THERE WAS SPARKING ON THE SCREEN. INCIDENT OCCURRED BEFORE THE PROCEDURE; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8537772 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page