ARIS COBLATION TURBINATE REDUCTION WAND 72290113Recalls & reported problems
Smith & Nephew, Inc. · ARIS COBLATION TURBINATE REDUCTION WAND · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00885556796047), received between Oct 9, 2024 and May 11, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Peeled/Delaminated3
- Detachment of Device or Device Component2
- Flaked2
- Melted2
- Degraded1
- Delivered as Unsterile Product1
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Flaked; Melted
IT WAS REPORTED THAT DURING A TURBINATE SURGERY, THE GLUE AND UNDERLYING BLACK SHEATHING OF ONE (1) ARIS TURBINATE REDUCTION WAND MELTED AWAY, THERE WAS FRAGMENTED GLUE OR BLACK INSULATION THAT CAN COME OFF. IT IS UNKNOW HOW THE PROCEDURE WAS PERFORMED OR IF THERE WAS ANY DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25141216 ↗MalfunctionMay 11, 2026Flaked; Melted
IT WAS REPORTED THAT DURING A TURBINATE SURGERY, THE GLUE AND UNDERLYING BLACK SHEATHING OF ONE (1) ARIS TURBINATE REDUCTION WAND MELTED AWAY, THERE WAS FRAGMENTED GLUE OR BLACK INSULATION THAT CAN COME OFF. IT IS UNKNOWN HOW THE PROCEDURE WAS PERFORMED OR IF THERE WAS ANY DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25141209 ↗MalfunctionMar 9, 2026Material Separation; Suction Problem
IT WAS REPORTED THAT, DURING A TURBINATE REDUCTION, AN ARIS WAND KEPT CLOGGING; STYLET, SALINE RINSE, AND GENTLE WIPING WITH WET 4X4 WERE USED, BUT THE DEVICE KEPT CLOGGING. THE DEVICE THEN WAS BACKFLUSHED WITH SALINE AND ABLATION PEDAL ACTIVATED. IT WAS ALSO NOTICED THAT THERE WAS A PEELING OF THE ELECTRODE, IT LOOKED LIKE TISSUE CHAR. THE SURGEON INSPECTED THE PEELING AND PULLED IT OFF, IT WAS A PORTION OF THE ELECTRODE, NOT CHAR. THE PROCEDURE WAS RESUMED, AFTER A 4-MIN SURGICAL DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547842 ↗MalfunctionFeb 10, 2026Detachment of Device or Device Component; Electrical Shorting
IT WAS REPORTED THAT DURING THE TURBINATE REDUCTION PROCEDURE, IT WAS NOTICED THAT THE ARIS TURBINATE REDUCTION WAND HAD A "WAND SHORT CIRCUIT" ERROR MESSAGE POPPED UP, AND THE ACTIVE ELECTRODE FELL OFF THE TIP OF THE ARIS WAND INTO THE PATIENT'S NASAL CAVITY. THE ELECTRODE WAS RETRIEVED USING GRASPING FORCEPS. THE PROCEDURE WAS PERFORMED WITHOUT ANY SURGICAL DELAY, USING THE SAME PRODUCT AS AFTER HITTING THE BACK BUTTON, IT WORKED, AND SURGEON PROCEEDED WITH THE CASE. NO PATIENT INJURIES, HARM OR FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24301706 ↗MalfunctionDec 3, 2025Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, A PIECE OF THE TURBINATE REDUCTION WAND ELECTRODE BROKE OFF THE FLAT ELECTRODE AND WAS PROTRUDING OUTWARDS, BUT IT WAS STILL CONNECTED. IT DID NOT DETACH IN THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23705297 ↗MalfunctionNov 20, 2025Break; Break
IT WAS REPORTED THAT, DURING AN UNKNOWN PROCEDURE, WHEN THEY ENDED USING THE ARIS TURBINATE REDUCTION WAND, THE TIP FELL OFF. NO FURTHER COMPLICATIONS WERE REPORTED. IT IS UNKNOWN IF THERE WAS A BACKUP DEVICE AVAILABLE OR IF THERE WAS A DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23605659 ↗MalfunctionSep 19, 2025Peeled/Delaminated; Manufacturing, Packaging or Shipping Problem; Peeled/Delaminated; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT, BEFORE A TURBINATE REDUCTION, WHEN THE ARIS WAND WAS TAKEN OUT OF THE PACKAGING, A WHITE CIRCLE WAS OBSERVED IN THE MIDDLE OF THE ELECTRODE. SURGEON DECIDED TO STILL USE THE DEVICE FOR THE SURGERY. DURING THE SECOND TURBINATE REDUCTION, IT WAS NOTICED THAT THE COATING AND THE WHITE CIRCLE HAD PEELED BACK. THE PROCEDURE WAS COMPLETED WITHOUT A SURGICAL DELAY USING THE SAME S+N PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23096030 ↗MalfunctionSep 19, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT, DURING A TURBINATE REDUCTION, IT WAS NOTICED THAT BLACK COATING THAT COVERS THE ARIS WAND PEELED OFF, NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT A SURGICAL DELAY USING AN EQUIVALENT SN BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23095999 ↗MalfunctionSep 18, 2025Degraded; Device Contaminated During Manufacture or Shipping; Degraded; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DURING AN ENT PROCEDURE, ONE (1) ARIS TURBINATE WAND HAD A BLACK, PLASTIC-LIKE MATERIAL. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23085935 ↗MalfunctionAug 18, 2025Break; Break
IT WAS REPORTED THAT DURING A SEPTOPLASTY THE ARIS TURBINATE REDUCTION WAND BROKE WHILE IT WAS BEING RETRACTED, THE SURGERY WAS COMPLETED USING A BACK-UP DEVICE, THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22825107 ↗MalfunctionJul 15, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING INSPECTION, IT WAS NOTICED THAT THE ARIS TURBINATE WAND FROM THE BOX WAS NOT SEALED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22513761 ↗MalfunctionMay 2, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT DURING SURGERY, AFTER THE ARIS TURBINATE WAND WAS REMOVED FROM THE TURBINATE, 15 SECONDS LATER, THE PLASTIC COATING ON THE SHAFT PEELED BACK. THE COATING ON THE SHAFT APPEARED TO BE STILL FULLY INTACT, AND IT APPEARED THAT NONE OF THE PLASTIC COATING REMAINED IN THE PATIENT. SURGERY WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21952615 ↗MalfunctionOct 24, 2024Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT DURING A TURBINATE REDUCTION, THE ARIS TURBINATE WAND WAS OVERHEATING AND EXPELLING AN EXCESSIVE AMOUNT OF SMOKE. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20527744 ↗MalfunctionOct 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LARGE BONY TURBINATE REDUCTION, THE ARIS TURBINATE REDUCTION WAND INSULATION ON THE INSTRUMENT SEPARATED FROM THE SHAFT. THE WEREWOLF MACHINE THEN HAD AN ERROR CODE AND THE WAND STOPPED WORKING. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A COMPETITOR DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20407183 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page