Skip to content

None 470384Recalls & reported problems

Intuitive Surgical, Inc. · BIPOLAR CAUTERY CORD · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00886874112601), received between May 27, 2023 and Sep 17, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Thermal Decomposition of Device4
  • Melted1
  • No Apparent Adverse Event1
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Thermal Decomposition of Device; Melted

    It was reported that during a da Vinci-assisted surgical procedure, the "code" on the Bipolar energy cable had melted. The procedure was completing as planned with no reported injury.Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: The customer reported no sparking, smoke, fire, or thermal damage. The cable was inspected prior to the event with no damage noted, and the number of prior uses and how long it had been on site were both unknown. At the time, the cable was connected to an Integrated Electro Surgical Unit (IESU) generator using bipolar energy and was plugged into Maryland and Fenestrated Bipolar instruments; no information was provided on whether it was connected to the correct generator port or whether a third-party generator was in use. There were no injuries, the issue was resolved by switching to a backup cable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50811422 ↗
  • MalfunctionJun 16, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT DURING DA VINCI-ASSISTED SURGICAL PROCEDURE, THE BIPOLAR ENERGY CORD WAS NOT WORKING. THE PROCEDURE WAS ABORTED/CONVERTED DUE TO ANATOMY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25521467 ↗
  • MalfunctionApr 29, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE BIPOLAR ENERGY CABLE WAS INVOLVED IN A SPARKING INCIDENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035629 ↗
  • MalfunctionJan 27, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT THE BIPOLAR ENERGY CABLE HAD A POOR CONNECTION AND SHOWED SIGNS OF BURNING AT THE CONNECTION SITE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192968 ↗
  • MalfunctionJan 23, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT THE BIPOLAR ENERGY CABLE HAD POOR CONNECTION AND BURNED IN CONNECTION. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24166280 ↗
  • MalfunctionMay 27, 2023No Device Output; No Device Output

    IT WAS REPORTED THAT DURING A VINCI-ASSISTED SURGICAL PROCEDURE, WHEN THE MONOPOLAR CURVED SCISSORS AND MARYLAND BIPOLAR FORCEPS INSTRUMENTS WERE INTRODUCED, AND THE MONOPOLAR AND BIPOLAR CABLES WERE CONNECTED, NO ENERGY WAS PASSED TO THE INSTRUMENTS. THE CABLES WERE REPLACED; HOWEVER, THE ISSUE PERSISTED. THE SURGEON MADE THE DECISION TO CONVERT THE PROCEDURE TO LAPAROSCOPIC SURGERY WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: UPON ENCOUNTERING THE ENERGY ISSUE, THE CUSTOMER RE-STARTED THE ERBE, CHECKED THE CABLES, AND EXCHANGED TO ANOTHER PAIR OF ACTIVATION CABLES; HOWEVER, THE ENERGY CONNECTION REMAINED UNSUCCESSFUL. AN ISI FIELD SERVICE ENGINEER (FSE) DURING A MAINTENANCE VISIT HAD ADVISED THE CUSTOMER TO ACQUIRE NEW ACTIVATION CABLES. A VISUAL INSPECTION WAS PERFORMED, AND DAMAGE WAS NOTED ON ONE CABLE. NO PATIENT HARM OR INJURY WAS REPORTED. THE SURGEON UNDOCKED FROM THE DA VINCI SYSTEM AND CONTINUED THE PROCEDURE LAPAROSCOPICALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17015837 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page