Skip to content

Arthrex® AR-9705A-90Recalls & reported problems

Arthrex, Inc. · OPES 90, ASPIRATING ABLATOR, TOOTHBRUSH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867062627), received between Aug 16, 2017 and Nov 7, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Component Missing1
  • Detachment of Device or Device Component1
  • Failure to Power Up1
  • Material Deformation1

Patient problems

  • Electric Shock1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2024Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THE PROBE COULD NOT BE ACTIVATED AFTER CONNECTION AND THE DEVICE WAS WITHOUT FUNCTION. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR D OA SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20635681 ↗
  • InjuryNov 14, 2019Break; Component Missing; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT A SHOULDER ARTHROSCOPY WAS PERFORMED WITH AN ARTHREX DEVICE. THE HEAD OF THE PROBE GOT DEFORMED AND DETACHED BY THE EFFECT OF HEAT. ACCORDING TO THE SURGEON THE DETACHED FRAGMENT WAS NOT RETRIEVED. NO FURTHER INFORMATION RECEIVED AT THE MOMENT. UPDATE 25-OCT-2019: AFTER THE SURGERY IT WAS FOUND THAT THE PROBE HAD A BROKEN EDGE ON THE HEAD AND A SMALL PIECE WAS MISSING, THE JOINT WAS WELL WASHED AND SEARCHED FOR THE MISSING PIECE, BUT WAS NOT FOUND.

    Patient problems: Foreign Body In Patient

    MAUDE report 9324242 ↗
  • InjuryAug 16, 2017Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON WAS ATTEMPTING TO USE AN OPES 90° TOOTHBRUSH ASPIRATING ABLATOR DURING USE, THE SURGEON EXPERIENCED AN ELECTRODE SHOCK IN HIS HAND THROUGH THE COOLCUT HAND PIECE. THE CASE WAS COMPLETED USING THE SAME DEVICE WITH NO FURTHER ISSUES.

    Patient problems: Electric Shock

    MAUDE report 6798113 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →