VULCAN 7210812FRecalls & reported problems
Smith & Nephew, Inc. · VULCAN GENERATOR ONLY REFURBISHED CE MARK · Class 2
All models of VULCAN (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03596010613561), received between Feb 26, 2019 and Dec 27, 2022 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver Energy” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Deliver Energy2
- Overheating of Device2
- Protective Measures Problem2
- Calibration Problem1
- Electrical Shorting1
- Insufficient Device Problem Information1
- Intermittent Loss of Power1
- Material Deformation1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions4
- No Information1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 27, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE VULCAN GENERATOR DISPLAYED A TEMPERATURE WARNING AND WAS OVERHEATING. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16056744 ↗MalfunctionApr 20, 2021Failure to Deliver Energy; Material Deformation; Failure to Deliver Energy; Material Deformation
IT WAS REPORTED THAT THE VULCAN GENERATOR WAS NOT WORKING. THIS WAS NOTICED DURING INSPECTION; THEREFORE, NO PATIENT WAS INVOLVED. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT DID NOT HAVE RF OUTPUT WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11694080 ↗MalfunctionMar 28, 2021Calibration Problem; Protective Measures Problem; Calibration Problem; Protective Measures Problem
IT WAS REPORTED THAT, DURING A WRIST ARTHROSCOPY USING A "VULCAN GENERATOR"; THE USER SAID THAT THEY WERE GETTING THE VOLTAGE CALIBRATION FAIL ALARM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11580381 ↗MalfunctionFeb 22, 2021Failure to Deliver Energy; Failure to Deliver Energy
IT WAS REPORTED THAT, DURING A HIP SCOPE, THE VULCAN GENERATOR WOULDN'T SEND POWER TO THE WAND. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING THE SAME DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11359249 ↗MalfunctionNov 9, 2020Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING SURGERY, THE VULCAN GENERATOR HAD A VOLTAGE CAL FAILURE ERROR AFTER IT WAS ON FOR A PERIOD OF TIME. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10813942 ↗InjurySep 18, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A PROCEDURE, THE GENERATOR WAS FAULTING AND THAT THE ATTACHED FOOT PEDAL HAD A DAMAGED CABLE, NO MORE INFORMATION. THERE WAS NO DELAY AND THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE.
Patient problems: No Information
MAUDE report 10556805 ↗MalfunctionAug 24, 2020Noise, Audible; Intermittent Loss of Power; Noise, Audible; Intermittent Loss of Power
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, THE VULCAN GENERATOR WAS MAKING A LOUD BEEPING NOISE AND SOMETIMES IT WOULD SHUT OFF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10446404 ↗MalfunctionDec 10, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE GENERATOR WAS OVERHEATING. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO DELAY AND NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9447403 ↗MalfunctionMay 7, 2019Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY THE GENERATOR WAS NOT WORKING. NO BACK UP DEVICE WAS AVAILABLE, NO DELAY NOR PATIENT INJURIES WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8586491 ↗MalfunctionFeb 26, 2019Electrical Shorting
IT WAS REPORTED THAT DURING A ROTATOR ARTHROSCOPY, THE DEVICE SHORTED OUT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8372516 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page