STEPHANIX D2RS_9090_B1_NV2Recalls & reported problems
Stephanix · D2RS is a direct digital dynamic remote-controlled fluoroscopy and radiography system equipped with the latest generation of Flat Panel Detector (FPD). The single FPD can perform both fluoroscopy and radiography. Not for mammography. · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03664049001052), received between Jul 15, 2025 and Jul 15, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unintended Movement” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2025Unintended Movement; Unintended Movement
DURING A MOVEMENT ORDERED FROM THE CONSOLE BY A RADIOLOGY OPERATOR, THE TABLE DID NOT STOP. NO PATIENT, NO INJURY. THE USER DID NOT USE THE EMERGENCY STOP BUTTONS. AFTER ANALYSIS OF THE LOGS BY OUR SUPPORT TEAM, WE FOUND THAT WHEN THE TILTING CONTROL WAS RELEASED, A BUG OCCURRED IN THE INVERTERS AND MOVEMENT CONTINUED WITHOUT REACHING THE LIMIT SWITCH. WE HAVE IMPLEMENTED A HARDWARE SAFETY WHICH MONITORS IF AN ABNORMAL SITUATION LIKE THIS OCCURS AND CUTS THE POWER SO THAT THE MOVEMENT STOPS IMMEDIATELY. DURING THIS ISSUE, THE CONTACTOR FAILED TO OPEN PROPERLY AND THE EQUIPMENT REMAINED POWERED. OUR TECHNICAL TEAM ADJUSTED THE LIMIT SWITCH AND CHANGED THE CONTACTOR (MANUFACTURED BY SCHNEIDER ELECTRIC), THE COMPONENT THAT CUTS OFF THE POWER WHEN THE PROBLEM OCCURS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22506256 ↗
Other System, X-Ray, Fluoroscopic, Image-Intensified devices
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified9 recalls7 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified7 recalls1 report
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified6 recalls4 reports
- ProxiDiagnost N90 706110Philips Medical Systems DMC GmbH · System, X-Ray, Fluoroscopic, Image-Intensified5 recalls1 report
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- CombiDiagnost R90 709031Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls11 reports
- ProxiDiagnost N90 706100Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls6 reports
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000CCanon Inc. · System, x-ray, fluoroscopic, image-intensified4 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.