Skip to content

STEPHANIX D2RS_9090_B1_NV2Recalls & reported problems

Stephanix · D2RS is a direct digital dynamic remote-controlled fluoroscopy and radiography system equipped with the latest generation of Flat Panel Detector (FPD). The single FPD can perform both fluoroscopy and radiography. Not for mammography. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03664049001052), received between Jul 15, 2025 and Jul 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2025Unintended Movement; Unintended Movement

    DURING A MOVEMENT ORDERED FROM THE CONSOLE BY A RADIOLOGY OPERATOR, THE TABLE DID NOT STOP. NO PATIENT, NO INJURY. THE USER DID NOT USE THE EMERGENCY STOP BUTTONS. AFTER ANALYSIS OF THE LOGS BY OUR SUPPORT TEAM, WE FOUND THAT WHEN THE TILTING CONTROL WAS RELEASED, A BUG OCCURRED IN THE INVERTERS AND MOVEMENT CONTINUED WITHOUT REACHING THE LIMIT SWITCH. WE HAVE IMPLEMENTED A HARDWARE SAFETY WHICH MONITORS IF AN ABNORMAL SITUATION LIKE THIS OCCURS AND CUTS THE POWER SO THAT THE MOVEMENT STOPS IMMEDIATELY. DURING THIS ISSUE, THE CONTACTOR FAILED TO OPEN PROPERLY AND THE EQUIPMENT REMAINED POWERED. OUR TECHNICAL TEAM ADJUSTED THE LIMIT SWITCH AND CHANGED THE CONTACTOR (MANUFACTURED BY SCHNEIDER ELECTRIC), THE COMPONENT THAT CUTS OFF THE POWER WHEN THE PROBLEM OCCURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22506256 ↗