AESCULAP PM433RRecalls & reported problems
Aesculap AG · ADTEC BIPOLAR Grasping Forceps, jaw inserts, straight, (310 mm), diam. 5 mm, fenestrated, double action, reusable · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04038653208654), received between Apr 14, 2021 and Nov 3, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Fragmentation1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PM433R-JAW INS.BIP.MACRO FORCEPS D:5/310MM. ACCORDING TO THE COMPLAINT DESCRIPTION, IT WAS REPORTED THAT DURING A HYSTERECTOMY SURGERY ON (B)(6) 2021 THE DOCTOR NOTICED THAT ALTHOUGH THE DEVICE PM403R MAKE THE USUAL SOUNDS AND PRODUCE WARM FROM THE BIPOLAR GENERATOR. THE FORCEPS ITSELF DOESN'T FUNCTION AS THEY SHOULD BE. AS THE PROXIMAL END BASE GOT WARM INSTEAD OF THE FORCEPS ITSELF. THE DOCTOR SWITCHED TO A ANOTHER CABLE (GN133) AND THE PROBLEM CONTINUED. THE DOCTOR DECIDED TO CONTINUE THE SURGERY WITHOUT THE PM403R. THE PATIENT HAS NO INJURIES AT ALL AND THEY MANAGED TO COMPLETE THE OPERATION SUCCESSFULLY. THERE WAS NO DESCRIBED PATIENT HARM. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED AS OF THIS REPORT. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4). INVOLVED COMPONENTS: PM973R-INSULATED OUTER TUBE 5/5MM 310MM-UNKNOWN. PM450R-HANDLE FOR BIPOLAR INSTRUMENTS-UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12749140 ↗MalfunctionApr 14, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED TO AESCULAP AG THAT A JAW INS.BIP.MACRO FORCEPS D:5/310MM (PART # PM433R) WAS USED DURING AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, THE CERAMIC CRACKED INSIDE THE GRASPING FORCEPS. FOLLOWING THE PROCEDURE, DURING THE PRE-STERILIZATION INSPECTION, THE CERAMIC CRACK INSIDE THE GRASPING FORCEPS WAS IDENTIFIED. NO POWER FAILURE OCCURRED DURING USE. THERE WERE NO NOTABLE DIFFICULT SCENES DURING THE OPERATION. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ALTHOUGH REQUEST, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11664838 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page