AESCULAP GK375RRecalls & reported problems
Aesculap AG · INNER TUBE W/HANDLE FOR GK374R · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04038653457847), received between Jun 4, 2021 and Jun 4, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2021Break; Melted; Break; Melted
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH GK375R-INNER TUBE W/HANDLE FOR GK374R. ACCORDING TO THE COMPLAINT DESCRIPTION, THE HOOK INSULATION FAILED DURING SURGERY. DURING A LAPAROSCOPIC PROCEDURE THERE WAS BLEEDING AND THE CURRENT CAUTERY SETTING WAS NOT EFFECTIVE IN CONTROLLING THE BLEEDING SO THE CAUTERY SETING WAS INCREASED. IT WAS NOT NOTICED UNTIL THE END OF THE CASE, WHEN THE INSTRUMENTS WERE DECONTAMINATED, THAT THE INSULATION APPEARED TO BE MELTED. THERE WAS NO DESCRIBED PATIENT HARM OR SURGICAL DELAY. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER (B)(4). ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00318 ((B)(4) GK383R).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11938838 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page