Olympus WA00013ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · HF-cable, 4 m, UES-40 HF-unit to TURis resectoscope · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04042761030489), received between Jun 3, 2021 and Nov 26, 2025 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Sparking2
- Activation Problem1
- Arcing1
- Electrical /Electronic Property Problem1
- Insufficient Device Problem Information1
- Labelling, Instructions for Use or Training Problem1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryNov 26, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 23660857 ↗MalfunctionOct 13, 2023Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE HF-CABLE HAD INSTRUCTIONS FOR USE THAT WERE POSSIBLY INCORRECT. THE EVENT WAS NOTICED PRIOR TO A THERAPEUTIC TRANSURETHRAL RESECTION IN SALINE PROCEDURE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17930271 ↗MalfunctionSep 22, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE HF-CABLE HAD AN ER08 AND AN OPEN CIRCUIT. THE EVENT OCCURRED DURING A PROCEDURE. THE PROCEDURE WAS FINISHED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS EVALUATED, AND IT WAS FOUND THAT THERE WAS A DAMAGED CABLE. THIS MDR (MEDICAL DEVICE REPORT) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17798781 ↗MalfunctionNov 15, 2022Sparking; Sparking
THE CUSTOMER REPORTED TO OLYMPUS DURING PREPARATION FOR A THERAPEUTIC PROCEDURE, SPARKS WERE PRESENT WHEN THE HF-CABLE WAS CONNECTED TO THE UES-40. UPON RECONNECTING, THE DEVICE DID NOT POWER ON. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15802271 ↗MalfunctionNov 26, 2021Break; Arcing; Sparking; Break; Arcing; Sparking
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION IN SALINE (TURIS) PROCEDURE, THE HF-CABLE BROKE AND SPARKED. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SIMILAR CABLE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12877772 ↗InjuryJun 3, 2021Output Problem; Activation Problem; Output Problem; Activation Problem
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC TRANSURETHRAL RESECTION IN SALINE (TURIS) PROCEDURE, ERROR MESSAGE E02 OCCURRED FREQUENTLY AND THE PROCEDURE COULD NOT BE COMPLETED BUT HAD TO BE POSTPONED. THE INTENDED PROCEDURE WAS THEN COMPLETED ON ANOTHER DAY WITH A DIFFERENT HF CABLE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11929036 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗