Olympus WA69334MRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Jaws insert "HICURA", 5 x 330, wave type, monopolar · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04042761079426), received between Sep 18, 2023 and Mar 10, 2026 — classified as 4 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Failure to Align4
- Detachment of Device or Device Component3
- Material Deformation2
- Mechanics Altered1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- Foreign Body In Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS, REUSABLE DEVICE HAD A BROKEN DISTAL END. THE ISSUE WAS OBSERVED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24560641 ↗MalfunctionDec 12, 2025Break; Break
IT WAS REPORTED THAT DURING SET UP FOR AN UNKNOWN PROCEDURE, THE TIP BROKE OFF THE INSERT ON ASSEMBLY. THE TIP WAS LOOSE AND HAD TO BE PINCHED TOGETHER TO GET IT INTO THE SHAFT. THERE WAS NO REPORTED PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23790108 ↗InjuryDec 12, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TIP BROKEN OFF IN THE PATIENT'S ABDOMEN. THERE WAS NO REPORTED INJURY OR HARM TO THE PATIENT. FOLLOW UP WITH THE CUSTOMER FOR MORE INFORMATION IS PENDING.
Patient problems: Foreign Body In Patient
MAUDE report 23789961 ↗MalfunctionDec 12, 2025Break; Break
NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23789953 ↗MalfunctionJul 14, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THE WORKING TIPS OF THE JAWS EASILY BECAME MISALIGNED RENDERING THE INSERT USELESS. THE ISSUE OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22495728 ↗InjuryJun 26, 2025Break; Break
IT WAS REPORTED DURING SEDATION FOR AN UNKNOWN THERAPEUTIC LAPAROSCOPIC PROCEDURE PART OF THE DEVICE BROKE OFF INTO THE ABDOMINAL AREA. THE PART WAS RETRIEVED AND THE PROCEDURE WAS COMPLETED USING A BACK UP DEVICE. NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338429 ↗InjuryJun 26, 2025Break; Break
IT WAS REPORTED DURING SEDATION FOR AN UNKNOWN THERAPEUTIC LAPAROSCOPIC PROCEDURE PART OF THE DEVICE BROKE OFF INTO THE ABDOMINAL AREA. THE PART WAS RETRIEVED AND THE PROCEDURE WAS COMPLETED USING A BACK UP DEVICE. NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22338338 ↗InjuryJun 26, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE WAVY GRASPER TIP BROKE OFF IN THE PERITONEAL SPACE OF THE PATIENT. IT WAS NOT NOTICED IMMEDIATELY. THE SCRUB NURSE LATER NOTICED THAT THE TIP WAS MISSING FROM THE INSTRUMENT. THE SURGEON WENT INTO THE PERITONEAL SPACE WITH ANOTHER INSTRUMENT, LOCATED THE WAVY GRASPER TIP AND REMOVED IT FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A BACKUP DEVICE. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22334269 ↗MalfunctionAug 26, 2024Failure to Align; Failure to Align
IT WAS REPORTED THE RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS JAWS WERE MISALIGNED. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20069291 ↗MalfunctionAug 22, 2024Failure to Align; Failure to Align
IT WAS REPORTED THAT THE RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS JAWS WERE MISALIGNED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20028893 ↗MalfunctionAug 19, 2024Break; Break
IT WAS REPORTED THAT THE RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS DISTAL END OF THE JAW WAS BROKEN OR DEFORMED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19999880 ↗MalfunctionOct 20, 2023Failure to Align; Failure to Align
THE CUSTOMER REPORTED WHEN USING THE DEVICE FOR THE FIRST TIME DURING A LAPAROSCOPIC LIVER RESECTION, IT WAS NOTICED THERE WAS SOME DISCOMFORT IN ITS OPERABILITY FROM THE BEGINNING, SO IT WAS CHECKED. THE CUSTOMER NOTICED THAT THE TIP OF THE JAWS WAS SLIGHTLY MISALIGNED. THE PROCEDURE WAS COMPLETED WITH THE SAME SET OF EQUIPMENT. NO HEALTH DAMAGE TO PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17979939 ↗MalfunctionSep 29, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT THE GRASPING FORCEPS HAD MISALIGNED JAWS AND COULD NOT BE USED. THE ISSUE WAS FOUND OUT OF THE BOX. NO PATIENT OR PROCEDURE WAS INVOLVED. THERE WERE NO REPORTS OF HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17848389 ↗MalfunctionSep 25, 2023Material Deformation; Material Deformation
THE CUSTOMER REPORTED THAT THE GRASPING FORCEPS HAD MISALIGNED JAWS AND COULD NOT BE USED. THERE WERE NO REPORTS OF HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17814079 ↗MalfunctionSep 18, 2023Failure to Align; Failure to Align
THE CUSTOMER REPORTED THAT HER OLYMPUS HICURA JAWS INSERT EXPERIENCED AN OUT OF BOX FAILURE. ACCORDING TO THE INITIAL REPORTER, THE GRASPING FORCEPS HAD MISALIGNED JAWS NOTED DURING THE INITIAL USE OF THE DEVICE. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17765033 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page