Olympus WA69416MRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04042761080002), received between Jan 26, 2022 and Mar 20, 2026 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Mechanical Problem3
- Sparking3
- Detachment of Device or Device Component1
- Fracture1
- Material Fragmentation1
- Material Split, Cut or Torn1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Foreign Body In Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Break
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE REUSABLE RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS HAD A BROKEN DISTAL COMPONENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654040 ↗MalfunctionJan 29, 2025Material Twisted/Bent; Material Twisted/Bent
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21266077 ↗InjuryMar 8, 2024Break; Melted; Detachment of Device or Device Component; Break; Melted; Detachment of Device or Device Component
IT WAS REPORTED, THAT THE PIN IN THE MOVABLE PART (FULCRUM PART) IS DAMAGED. AND SOME PARTS HAVE FALLEN OFF (SOME OF THE RECOVERED PIECES ARE PLACED IN A POUCH AND ATTACHED).
Patient problems: Foreign Body In Patient
MAUDE report 18864687 ↗MalfunctionJan 16, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE JAWS INSERT "HICURA", 5 X 330, MARYLAND, BIPOLAR HAD THE TIP SCORCHED DURING CLEANING WITH NO POWER BEING GENERATED ANYMORE THAN NECESSARY. THE ISSUE WAS FOUND DURING THE LAPAROSCOPIC UTERINE FIBROID ENUCLEATION PROCEDURE WHICH WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18521745 ↗MalfunctionNov 16, 2023Break; Sparking; Break; Sparking
THE CUSTOMER REPORTED TO OLYMPUS, THE JAWS INSERT "HICURA", 5 X 330, MARYLAND, BIPOLAR HAD THE INSULATION AT THE TIP OF THE FORCEPS DAMAGED. THE ISSUE WAS FOUND DURING AN UNKNOWN THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. EVIDENCE OF SPARKING WAS FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18150407 ↗MalfunctionSep 5, 2023Mechanical Problem; Material Split, Cut or Torn; Mechanical Problem; Material Split, Cut or Torn
THE CUSTOMER REPORTED TO OLYMPUS THAT THE JAWS INSERT "HICURA", 5 X 330, MARYLAND, BIPOLAR INNER PART TO MARYLAND BIPOLAR FORCEPS WAS BENT. THE EVENT WAS FOUND DURING REPROCESSING FOR A LAPAROSCOPIC HYSTERECTOMY. THERE WAS NO DELAY, AND THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS EVALUATED, AND IT WAS FOUND THAT THE WHITE COMPONENT/PART ABOVE THE JAWS APPEARED TO BE BROKEN OR TORN AND THE JAWS WOULD NOT CLOSE COMPLETELY. THIS MDR (MEDICAL DEVICE REPORT) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17681839 ↗MalfunctionAug 8, 2023Break; Break
THE CUSTOMER REPORTED THAT DURING AN UNSPECIFIED PROCEDURE THE TIP JAW WAS DAMAGED AND FELL OFF. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THE TIP THAT FELL OFF HAS BEEN RECOVERED AND THERE IS NO HEALTH HAZARD. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17482420 ↗MalfunctionAug 8, 2023Break; Break
AN OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER SPARKS EMITTED FROM THE TIP OF JAWS INSERT "HICURA", 5 X 330, MARYLAND, BIPOLAR (OUTPUT 20W). THE MALFUNCTION WAS FOUND DURING THE THERAPEUTIC PROCEDURE, A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED, AND OLYMPUS REPAIR CENTER IDENTIFIED THE DISTAL END OF THE JAW WAS BROKEN. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE EVALUATION OF THE RETURNED DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17481014 ↗MalfunctionJul 20, 2023Break; Mechanical Problem; Break; Mechanical Problem
THE CUSTOMER REPORTED TO OLYMPUS, JAWS INSERT HICURA, 5 X 330, MARYLAND, BIPOLAR HAD BEEN IN CIRCULATION FOR ROUGHLY A MONTH. WHILE THE FORCEPS WAS BEING DECONTAMINATED, IT WAS NOTICED THAT A WHITE PIECE WAS MISSING OR BURNT AT THE BASE OF THE JAW OF THE INSERT. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17358986 ↗MalfunctionMay 3, 2023Sparking; Sparking
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER THAT SPARKS FLEW FROM THE TIP OF THE JAWS INSERT "HICURA", 5 X 330, MARYLAND, BIPOLAR DURING A LAPAROSCOPIC HYSTERECTOMY. THE DEVICE WAS USED WITH A NON-OLYMPUS ELECTROSURGICAL UNIT WITH A 30 WATTS SETTING. THE USE OF THE DEVICE WAS DISCONTINUED. THE PROCEDURE WAS COMPLETED. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE INSULATION AND CERAMIC PIN WERE DAMAGED. THIS REPORT IS BEING SUBMITTED FOR THE EVENT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16865711 ↗MalfunctionFeb 9, 2023Sparking; Sparking
DURING SURGERY (LAPAROSCOPIC HYSTERECTOMY), SPARKS WERE GENERATED FROM THE TIP. THERE WAS NO FALLING OFF OF FRAGMENTS, ETC., AND THERE WAS NO HEALTH HAZARD. IT WAS REPLACED WITH ANOTHER ONE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16343393 ↗MalfunctionJan 26, 2022Break; Fracture; Material Fragmentation; Mechanical Problem; Break; Fracture; Material Fragmentation; Mechanical Problem
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC LAPAROSCOPIC HYSTERECTOMY PROCEDURE, THE SUREGON FELT THAT CAUTERIZING IN BIPOLAR BECAME DIFFICULT. THE DEPOSITS A THE DISTAL END OF THE JAWS INSERT WERE REMOVED WITH A GAUZE BUT CAUTERIZATION WAS STILL NOT POSSIBLE AND WHEN THE INSTRUMENT WAS REMOVED AGAIN TO BE REPLACED, A CRACK WAS NOTICED IN THE INSULATION AT THE DISTAL END OF THE JAWS. IT IS UNCLEAR WHETHER ANY FRAGMENTS MAY HAVE BROKEN OFF AND FALLEN INTO THE PATIENT. HOWEVER, THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER SIMILAR DEVICE AND THERE WAS NO REPORT ABOUT AN ADVERSE EVENT OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13357861 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page