Olympus WA90002ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04042761082969), received between Aug 24, 2022 and Sep 13, 2022 — classified as 6 injuries.
The most frequently reported specific device problem is “Output above Specifications” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Output above Specifications3
- Use of Device Problem3
Patient problems
- Necrosis6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 13, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002
Patient problems: Necrosis
MAUDE report 15414577 ↗InjurySep 13, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002
Patient problems: Necrosis
MAUDE report 15414562 ↗InjurySep 13, 2022Output above Specifications
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002
Patient problems: Necrosis
MAUDE report 15414502 ↗InjurySep 13, 2022Output above Specifications
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: (B)(4), SERIAL (B)(4), LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, (B)(4), CASE WITH PATIE
Patient problems: Necrosis
MAUDE report 15414491 ↗InjuryAug 24, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND TISSUE WAS EXPERIENCING NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE OLYMPUS REPRESENTATIVE PLANS TO PROVIDE THE CUSTOMER WITH AN ADDITIONAL IN-SERVICE ON OPERATING THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT.
Patient problems: Necrosis
MAUDE report 15288709 ↗InjuryAug 24, 2022Output above Specifications
IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND TISSUE WAS EXPERIENCING NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE OLYMPUS REPRESENTATIVE PLANS TO PROVIDE THE CUSTOMER WITH AN ADDITIONAL IN-SERVICE ON OPERATING THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT.
Patient problems: Necrosis
MAUDE report 15288646 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page