Skip to content

Olympus WA69424MRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Jaws insert "HICURA", 5 x 330, Precision grasper, bipolar · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04042761083669), received between Apr 17, 2025 and Apr 17, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Melted” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Melted3
  • Sparking3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Melted; Sparking; Melted; Sparking

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21859699 ↗
  • MalfunctionApr 17, 2025Melted; Sparking; Melted; Sparking

    IT WAS REPORTED THAT DURING THE PROCEDURE, WHILE USING THE SUBJECT DEVICE, WHEN OUTPUTTING IN A CLOSE STATE WITHOUT CONTACT WITH THE METAL OF THE MONOPOLAR MACHINE, SPARKS FLEW AND THE TIP OF THE SUBJECT DEVICE MELTED. THE THERAPEUTIC UTERINE MYOMECTOMY PROCEDURE WAS COMPLETED AFTER REPLACEMENT WITH A SIMILAR OLYMPUS BACK UP NEW DEVICE. THE PATIENT WAS UNDER SEDATION WITH DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS PROLONGED FOR ONLY A FEW MINUTES FOR THE REPLACEMENT OF THE DEVICE. PER ADDITIONAL INFORMATION MALFUNCTION OCCURRED ONLY DURING CUTTING OUTPUT, DID NOT OCCUR DURING COAG-OUTPUT. THERE WAS NO HEALTH HAZARD TO THE PATIENT OR THE DEVICE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21858938 ↗
  • MalfunctionApr 17, 2025Melted; Sparking; Melted; Sparking

    NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21857537 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates