Olympus WA99510ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Cable for neutral electrode · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04042761085892), received between Jul 28, 2022 and Nov 7, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “No Device Output” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- No Device Output1
- Use of Device Problem1
Patient problems
- Burn(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2023No Device Output; No Device Output
THE CUSTOMER REPORTED THAT DURING A POLYPECTOMY THE BLUE PEDAL AND GENERATOR DID NOT WORK. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE AND NO DELAY WAS REPORTED. THERE WERE NO REPORTS OF HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18088295 ↗MalfunctionJul 28, 2022Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE ADULT PATIENT HAD A SUNBURN-LIKE BURN (REDNESS) TO AN UNSPECIFIED LOCATION OF THE LEG WHERE THE SUBJECT DEVICE WAS PLACED, IN A STRAIGHT LINE. THE OLYMPUS REPRESENTATIVE SURMISES THE SUBJECT DEVICE WAS NOT PLACED PROPERLY ON THE PATIENT AND CAUSED A HEAT BUILDUP. THE SUBJECT DEVICE WAS NOT BURNT. THE THERAPEUTIC ANAL FISTULA PROCEDURE WAS COMPLETED SUCCESSFULLY. THE SUBJECT DEVICE WAS CONNECTED TO AN NON-OLYMPUS GENERATOR WITH THE SETTINGS AT 50/50 BLEND COAGULATION. THE SETTINGS WERE HIGH AND THE NON-OLYMPUS TEFLON PAD MELTED TOO. NO ALARMS WENT OFF DURING THE PROCEDURE.
Patient problems: Burn(s)
MAUDE report 15126995 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page