The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04042761085892), received between Jul 28, 2022 and Nov 7, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “No Device Output” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Device Output1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2023No Device Output; No Device Output

    THE CUSTOMER REPORTED THAT DURING A POLYPECTOMY THE BLUE PEDAL AND GENERATOR DID NOT WORK. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE AND NO DELAY WAS REPORTED. THERE WERE NO REPORTS OF HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18088295 ↗
  • MalfunctionJul 28, 2022Use of Device Problem; Use of Device Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THE ADULT PATIENT HAD A SUNBURN-LIKE BURN (REDNESS) TO AN UNSPECIFIED LOCATION OF THE LEG WHERE THE SUBJECT DEVICE WAS PLACED, IN A STRAIGHT LINE. THE OLYMPUS REPRESENTATIVE SURMISES THE SUBJECT DEVICE WAS NOT PLACED PROPERLY ON THE PATIENT AND CAUSED A HEAT BUILDUP. THE SUBJECT DEVICE WAS NOT BURNT. THE THERAPEUTIC ANAL FISTULA PROCEDURE WAS COMPLETED SUCCESSFULLY. THE SUBJECT DEVICE WAS CONNECTED TO AN NON-OLYMPUS GENERATOR WITH THE SETTINGS AT 50/50 BLEND COAGULATION. THE SETTINGS WERE HIGH AND THE NON-OLYMPUS TEFLON PAD MELTED TOO. NO ALARMS WENT OFF DURING THE PROCEDURE.

    Patient problems: Burn(s)

    MAUDE report 15126995 ↗