OLYMPUS Winter & Ibe GmbH · Electrosurgical generator "ESG-410" · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
8
Recalls naming it
2

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04042761086349), received between Oct 5, 2023 and May 22, 2026 — classified as 8 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 25, 2025Open, Classified

    Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

    Z-2339-2025 · Root cause: Component change control · FDA record ↗

  2. Jun 21, 2024Open, Classified

    Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery

    Z-2415-2024 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Failure to Deliver Energy2
  • Labelling, Instructions for Use or Training Problem2
  • Output Problem2
  • Defective Component1
  • Flare or Flash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2026Flare or Flash

    IT WAS REPORTED THAT DURING A TRANSURETHRAL RESECTION OF BLADDER TUMOR (TURBT) PROCEDURE THE MONOPOLAR CORD BEGAN SMOKING AND CAUGHT FIRE WHILE BEING PLUGGED IN TO THE PORT ON THE GENERATOR. NO HARM REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25255542 ↗
  • InjuryAug 15, 2025Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF A BLADDER TUMOR (TURBT) PROCEDURE, THE ELECTROSURGICAL GENERATOR HAD NO ENERGY. BOTH THE WIRED FOOTSWITCH AND WIRELESS FOOTSWITCH WERE ATTEMPTED. THE FOOTSWITCH INDICATOR ON THE DEVICE WAS INDICATING THE FOOTSWITCHES WERE PASSING A SIGNAL BUT THERE WAS NO ENERGY PRESENT. THERE WERE NO ERROR CODES PRESENT. THE PATIENT WAS UNDER GENERAL ANESTHESIA AT THAT TIME. THE CASE WAS SUBSEQUENTLY CANCELLED DUE TO THE EVENT. THE CUSTOMER RECEIVED A REPLACEMENT ELECTROSURGICAL GENERATOR. THE PATIENT WAS THEN BROUGHT BACK FOR A SECOND ATTEMPT PROCEDURE ON (B)(6) 2025 IN WHICH THEY UNDERWENT GENERAL ANESTHESIA AGAIN. THE SECOND PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO REPORTS OF PATIENT HARM. IT WAS REPORTED THAT NO TESTING OF THE DEVICE WAS DONE BEFORE THE FIRST PROCEDURE ATTEMPT. THE CUSTOMER PLUGGED THINGS IN AS USUAL AND BEGAN THE CASE. ALL APPEARED NORMAL UNTIL THE PHYSICIAN WAS IN THE PATIENT AND THE ELECTROSURGICAL GENERATOR WOULD NOT FIRE. IT DID NOT RESPOND TO PEDAL STEPS OR RESETS. IT DID NOT RESPOND TO NEW LOOPS. ALL CONNECTIONS WERE CHECKED AND RESET WITHOUT SUCCESS. THUS, THE CASE WAS CANCELLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22819404 ↗
  • InjuryAug 15, 2025Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED DURING A THERAPEUTIC TRANSURETHRAL RESECTION OF A BLADDER TUMOR (TURBT) PROCEDURE, THE ELECTROSURGICAL GENERATOR HAD NO ENERGY. BOTH THE WIRED FOOTSWITCH AND WIRELESS FOOTSWITCH WERE ATTEMPTED. THE FOOTSWITCH INDICATOR ON THE DEVICE WAS INDICATING THE FOOTSWITCHES WERE PASSING A SIGNAL BUT THERE WAS NO ENERGY PRESENT. THERE WERE NO ERROR CODES PRESENT. THE PATIENT WAS UNDER GENERAL ANESTHESIA AT THAT TIME. THE CASE WAS SUBSEQUENTLY CANCELLED DUE TO THE EVENT. THE CUSTOMER RECEIVED A REPLACEMENT ELECTROSURGICAL GENERATOR. THE PATIENT WAS THEN BROUGHT BACK FOR A SECOND ATTEMPT PROCEDURE ON (B)(6) 2025, IN WHICH THEY UNDERWENT GENERAL ANESTHESIA AGAIN. THE SECOND PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO REPORTS OF PATIENT HARM. IT WAS REPORTED THAT NO TESTING OF THE DEVICE WAS DONE BEFORE THE FIRST PROCEDURE ATTEMPT. THE CUSTOMER PLUGGED THINGS IN AS USUAL AND BEGAN THE CASE. ALL APPEARED NORMAL UNTIL THE PHYSICIAN WAS IN THE PATIENT AND THE ELECTROSURGICAL GENERATOR WOULD NOT FIRE. IT DID NOT RESPOND TO PEDAL STEPS OR RESETS. IT DID NOT RESPOND TO NEW LOOPS. ALL CONNECTIONS WERE CHECKED AND RESET WITHOUT SUCCESS. THUS, THE CASE WAS CANCELLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22819402 ↗
  • InjuryFeb 11, 2025Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    IT WAS REPORTED DURING SEDATION OF A THERAPEUTIC CYSTOSCOPY THE GENERATOR HAD THE INCORRECT QUICK REFERENCE GUIDE. THIS CAUSED A SURGICAL PROLONGATION GREATER THAN 30 MINUTES AND THE PROCEDURE WAS COMPLETED USING A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356076 ↗
  • InjuryFeb 11, 2025Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    IT WAS REPORTED DURING SEDATION OF A THERAPEUTIC CYSTOSCOPY THE GENERATOR HAD THE INCORRECT QUICK REFERENCE GUIDE. THIS CAUSED A SURGICAL PROLONGATION GREATER THAN 30 MINUTES AND THE PROCEDURE WAS COMPLETED USING A COMPETITOR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356065 ↗
  • InjuryNov 20, 2024Output Problem; Output Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20732910 ↗
  • InjuryNov 20, 2024Output Problem; Output Problem

    DURING SEDATION OF AN UNKNOWN THERAPEUTIC PROCEDURE THERE WAS A DELAY GREATER THAN 30 MINUTES DUE TO THE GENERATOR DISPLAYING AN E0728 INTERNAL HARDWARE FAILURE ERROR. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20732901 ↗
  • MalfunctionJun 7, 2024Defective Component; Defective Component

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE GENERATOR HAD A BAD FLEXRAY FIELD PROGRAMMABLE GATE ARRAY BOARD WHICH CAUSED THE ERROR 1010. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19485182 ↗
  • InjuryOct 5, 2023Communication or Transmission Problem; Communication or Transmission Problem

    AN OLYMPUS REPRESENTATIVE REPORTED, ON BEHALF OF THE CUSTOMER, THERE WAS AN ERROR CODE 1010 DISPLAYED AND THE DEVICE WOULD SHUT DOWN WHEN USING AN ELECTROSURGICAL GENERATOR "ESG-410". THE EVENT OCCURRED DURING THE THERAPEUTIC PROCEDURE (TRANSURETHRAL RESECTION OF BLADDER TUMOR-TURBT) WHICH LED TO THE CASE BEING ABORTED. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17881621 ↗
  • InjuryOct 5, 2023Communication or Transmission Problem; Communication or Transmission Problem

    AN OLYMPUS REPRESENTATIVE REPORTED, ON BEHALF OF THE CUSTOMER, THERE WAS AN ERROR CODE 1010 DISPLAYED AND THE DEVICE WOULD SHUT DOWN WHEN USING AN ELECTROSURGICAL GENERATOR "ESG-410". THE EVENT OCCURRED DURING THE THERAPEUTIC PROCEDURE (TRANSURETHRAL RESECTION OF BLADDER TUMOR-TURBT) WHICH LED TO THE CASE BEING ABORTED. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17881541 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →