OLYMPUS Winter & Ibe GmbH · Electrosurgical generator ESG-410 · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04042761087698), received between Feb 11, 2025 and May 15, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 25, 2025Open, Classified

    Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.

    Z-2338-2025 · Root cause: Component change control · FDA record ↗

  2. Sep 4, 2024Open, Classified

    There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).

    Z-0264-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Activation Failure2
  • Circuit Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Circuit Failure; Circuit Failure

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GENERATOR EXHIBITED THE HIGH VOLTAGE POWER SUPPLY BOARD HAD BLOWN CAPACITOR. AFTER REPLACING THE HIGH VOLTAGE POWER SUPPLY BOARD AND FUSES THE UNIT FAILED SELF-TEST WITH AN ERROR 0432. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22023844 ↗
  • InjuryFeb 11, 2025Activation Failure; Activation Failure

    IT WAS REPORTED WHEN USING THE GENERATOR WITH A THUNDERBEAT HANDPIECE, THERE WAS NO ACTIVATION. THE ISSUE OCCURRED DURING A SLEEVE GASTRECTOMY PROCEDURE WHILE THE PATIENT WAS SEDATED. THE USER CONNECTED A BOVIE PENCIL AND BOVIE PAD AND HAD THE SAME ISSUE. THE GENERATOR GAVE AN E0728 SYSTEM ERROR AND AN E1809 NEUTRAL ELECTRODE DISCONNECTED ERROR. THE USER TRIED ANOTHER PAD AND ELECTRODE, BUT THE DEVICE CONTINUED TO ERROR. THE PROCEDURE WAS COMPLETED WITH MANUAL OVER SEWING AND WAS PROLONGED GREATER THAN 30 MINUTES. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21359549 ↗
  • InjuryFeb 11, 2025Activation Failure; Activation Failure

    IT WAS REPORTED WHEN USING THE GENERATOR WITH A THUNDERBEAT HANDPIECE, THERE WAS NO ACTIVATION. THE ISSUE OCCURRED DURING A SLEEVE GASTRECTOMY PROCEDURE WHILE THE PATIENT WAS SEDATED. THE USER CONNECTED A BOVIE PENCIL AND BOVIE PAD AND HAD THE SAME ISSUE. THE GENERATOR GAVE AN E0728 SYSTEM ERROR AND AN E1809 NEUTRAL ELECTRODE DISCONNECTED ERROR. THE USER TRIED ANOTHER PAD AND ELECTRODE, BUT THE DEVICE CONTINUED TO ERROR. THE PROCEDURE WAS COMPLETED WITH MANUAL OVER SEWING AND WAS PROLONGED GREATER THAN 30 MINUTES. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21359460 ↗