AESCULAP PL740SURecalls & reported problems
Aesculap AG · CAIMAN DISP.INSTR.NON ARTICUL.D5/360MM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04046964693960), received between Dec 1, 2020 and Feb 18, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Arcing of Electrodes” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Arcing of Electrodes1
- Break1
- Energy Spectrum Incorrect1
- Material Fragmentation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2022Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED TO AESCULAP AG THAT A CAIMAN DISP.INSTR.NON ARTICUL.D5/360MM (PART # PL740SU) WAS USED DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CAIMAN PL470SU DEVICE BROKE INSIDE THE PATIENT. A X-RAY WAS PERFORMED TO RETRIEVE THE BROKEN JAW WHICH REPORTEDLY ADDED TWO (2) TO THREE (3) HOURS TO THE OPERATION. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT HARM WAS REPORTED AS A RESULT OF THE EVENT. AN ADDITIONAL MEDICAL INTERVENTION WAS NECESSARY. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13563081 ↗InjuryDec 29, 2021Energy Spectrum Incorrect; Energy Spectrum Incorrect
IT WAS REPORTED TO AECULAP AG THAT A CAIMAN EINM.INSTR.N.ABWINKELB.D5/360MM (PART # PL740SU) WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE CAIMAN USED IN THE PROCEDURE SOMETIMES FAILED TO COAGULATE OR ONLY PARTIALLY COAGULATED, DESPITE THE MACHINE WARNING THAT THE SYNTHESIS CYCLE HAD BEEN COMPLETED. IN PARTICULAR, WHEN THE USER OF THE PRODUCT CONTACTED, HE DESCRIBED THREE SITUATIONS WITHIN THE SURGERY WHERE THE PRODUCT CAUSED BLEEDING. THE FIRST, MORE SERIOUS ONE, THE MACHINE WARNED THAT THE SYNTHESIS/COAGULATION CYCLE HAD BEEN CARRIED OUT ON THE TISSUE IN THE BITE, BUT IN REALITY THERE WAS NO CURRENT DELIVERY, THIS MISLED THE DOCTOR WHO CUT THE TISSUE WITH THE APPROPRIATE CUTTING FUNCTION, WHICH GENERATED A BLEED THAT REQUIRED SUTURING. THE OTHER TWO WERE MINOR BLEEDS BECAUSE THERE WAS A PARTIAL CURRENT DELIVERED BY THE UNIT. THE COMPLAINT DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE ADVERSE EVENT / MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13112113 ↗MalfunctionMar 2, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PL740SU CAIMAN DISP.INSTR.NON ARTICUL.D5/360MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THE DEVICE BROKE DOWN, PROBLEMS DURING CUTTING OCCURED. THERE WAS NO DESCRIBED PATIENT HARM. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11399166 ↗MalfunctionDec 1, 2020Arcing of Electrodes; Arcing of Electrodes
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH PL740SU - CAIMAN DISP.INSTR.NON ARTICUL.D5/360MM. ACCORDING TO THE COMPLAINT DESCRIPTION, THERE WAS SPARKS ON THE JAW DURING USE (INSIDE THE ABDOMEN AND OUTSIDE). ERROR MESSAGE ON THE GENERATOR. ADDITIONAL INFORMATION WAS NOT PROVIDED NOR AVAILABLE / WAS NOT AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10924685 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page