Skip to content

n.a. 30675NDRecalls & reported problems

Karl Storz GmbH & Co. KG · MANHES High Frequency Needle · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04048551123799), received between Oct 24, 2018 and Apr 2, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Melted” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Melted2
  • Thermal Decomposition of Device2
  • Break1
  • Material Fragmentation1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Burn, Thermal1
  • Injury1
  • Tissue Damage1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2025Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted

    IT WAS REPORTED THAT THE INSULATION MELTED AT THE DISTAL END. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751592 ↗
  • MalfunctionApr 1, 2025Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted

    IT WAS REPORTED THAT THE INSULATION MELTED AT THE DISTAL END. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21748461 ↗
  • InjuryOct 24, 2018Break; Material Fragmentation; Smoking

    ALLEGEDLY, AS PER MEDWATCH #MW5080396 RECEIVED FROM THE FDA, DURING A PROCEDURE IN WHICH A MANHES HIGH FREQUENCY NEEDLE WAS BEING USED, WHEN THE CAUTERY WAS USED, THE SURGEON NOTED A SMALL AMOUNT OF SMOKE AND A FEW SMALL BLACK FRAGMENTS. THE SURGEON REMOVED THE DEVICE AND THEY NOTED A SMALL AREA ON THE INSTRUMENT WHERE THE DEVICE HAD A BREAK IN THE INSULATION APPROXIMATELY 1 CM FROM THE TIP. THE SURGEON THEN EXAMINED THE PATIENT LAPAROSCOPICALLY AGAIN AND NOTED A THROUGH AND THROUGH BURN TO THE PATIENT'S LEFT FALLOPIAN TUBE.

    Patient problems: Burn, Thermal; Injury; Tissue Damage

    MAUDE report 8002219 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page