ROBI 38151Recalls & reported problems
Karl Storz GmbH & Co. KG · RoBi® Plastic Handle · Class 2
All models of ROBI (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04048551325209), received between May 3, 2023 and Apr 3, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Energy Output Problem2
- Temperature Problem2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Display or Visual Feedback Problem1
- Electrical /Electronic Property Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT "BIPOLAR VIDEO LAPAROSCOPY FORCEPS DO NOT CONDUCT ELECTRICITY AND DO NOT COAGULATE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2026-00748. 9610617-2026-00755. 9610617-2026-00756.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24783406 ↗MalfunctionApr 3, 2026Energy Output Problem
IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781542 ↗MalfunctionApr 3, 2026Energy Output Problem
IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED; STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781541 ↗MalfunctionApr 23, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS A DEFECT. THE INTERNAL ASSESSMENT REVEALED THAT THE HANDLE IS BROKEN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: Insufficient Information
MAUDE report 21893613 ↗MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY, ITEM SPARKED INSIDE OF PATIENT. THERE WAS NO ONE FROM KARL STORZ TO WITNESSED THIS. MANAGER, ONSITE ENDOSCOPY BELIEVES IT IS THE INSERT, AND WOULD LIKE ALL THREE ITEMS EVALUATED TOGETHER. FEMALE PATIENT NOT HARMED, CASE WAS NOT DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19912420 ↗MalfunctionJul 5, 2023Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT 38700 ROBI METAL OUTER SHEATH. ACCORDING TO THE INFORMATION RECEIVED THE DOCTOR STATED THAT AT THE BEGINNING OF THE PROCEDURE THE INSTRUMENTS MELTED AND BURNED WHEN THEY WERE USED WITH THE OLDER EQUIPMENT. THERE WAS NO PATIENT/USERT HARM/INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17262399 ↗MalfunctionMay 3, 2023Temperature Problem; Temperature Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT 38151 ROBI PLASTIC HANDLE. ACCORDING TO THE INFORMATION RECEIVED THE DOCTOR WAS SHOCKED AND BURNED. THERE WAS NO PATIENT HARM/INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16859939 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page