Skip to content

ROBI 38151Recalls & reported problems

Karl Storz GmbH & Co. KG · RoBi® Plastic Handle · Class 2

All models of ROBI (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04048551325209), received between May 3, 2023 and Apr 3, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Energy Output Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Energy Output Problem2
  • Temperature Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Display or Visual Feedback Problem1
  • Electrical /Electronic Property Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT "BIPOLAR VIDEO LAPAROSCOPY FORCEPS DO NOT CONDUCT ELECTRICITY AND DO NOT COAGULATE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2026-00748. 9610617-2026-00755. 9610617-2026-00756.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24783406 ↗
  • MalfunctionApr 3, 2026Energy Output Problem

    IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781542 ↗
  • MalfunctionApr 3, 2026Energy Output Problem

    IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED; STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781541 ↗
  • MalfunctionApr 23, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS A DEFECT. THE INTERNAL ASSESSMENT REVEALED THAT THE HANDLE IS BROKEN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21893613 ↗
  • MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY, ITEM SPARKED INSIDE OF PATIENT. THERE WAS NO ONE FROM KARL STORZ TO WITNESSED THIS. MANAGER, ONSITE ENDOSCOPY BELIEVES IT IS THE INSERT, AND WOULD LIKE ALL THREE ITEMS EVALUATED TOGETHER. FEMALE PATIENT NOT HARMED, CASE WAS NOT DELAYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19912420 ↗
  • MalfunctionJul 5, 2023Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT 38700 ROBI METAL OUTER SHEATH. ACCORDING TO THE INFORMATION RECEIVED THE DOCTOR STATED THAT AT THE BEGINNING OF THE PROCEDURE THE INSTRUMENTS MELTED AND BURNED WHEN THEY WERE USED WITH THE OLDER EQUIPMENT. THERE WAS NO PATIENT/USERT HARM/INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17262399 ↗
  • MalfunctionMay 3, 2023Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT 38151 ROBI PLASTIC HANDLE. ACCORDING TO THE INFORMATION RECEIVED THE DOCTOR WAS SHOCKED AND BURNED. THERE WAS NO PATIENT HARM/INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16859939 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page