ROBI 38600Recalls & reported problems
Karl Storz GmbH & Co. KG · RoBi® Metal Outer Sheath · Class 2
All models of ROBI (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04048551328606), received between Nov 28, 2020 and Apr 3, 2026 — classified as 2 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Energy Output Problem3
- Smoking2
- Sparking2
- Use of Device Problem2
- Break1
- Display or Visual Feedback Problem1
- Electrical /Electronic Property Problem1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Foreign Body In Patient1
- Insufficient Information1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2026Energy Output Problem
IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781557 ↗MalfunctionApr 3, 2026Energy Output Problem
IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781539 ↗MalfunctionApr 3, 2026Energy Output Problem
IT WAS REPORTED THAT THE FORCEPS ARE NON-FUNCTIONAL, NO COAGULATION. THEY WERE DISASSEMBLED AND REASSEMBLED STILL NOT WORKING. AFTER REPLACING THE FORCEPS, THE SYSTEM WORKS WITHOUT ANY ISSUE, WITH NO CABLE CHANGE. THE PROCEDURE WAS COMPLETED AFTER REPLACING WITH FEW MINUTES DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781538 ↗MalfunctionApr 3, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT "BIPOLAR VIDEO LAPAROSCOPY FORCEPS DO NOT CONDUCT ELECTRICITY AND DO NOT COAGULATE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2026-00748, 9610617-2026-00755, 9610617-2026-00757.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24780584 ↗MalfunctionMar 13, 2026Smoking
IT WAS REPORTED THAT DURING ACTIVATION OF THE STORTZ BIPOLAR FORCEPS, SMOKE WAS RELEASED AT THE JOINT OF THE INSERT JAWS. ABSENCE OF COAGULATION AT THE JAWS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24596045 ↗MalfunctionFeb 26, 2026Break
THE CUSTOMER HAS NOT PROVIDED ANY INFORMATION ABOUT THE DEFECT OR MALFUNCTION. ACCORDING TO THE INVESTIGATION RESULTS: DISTAL LOOSENED CONNECTION. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2026-00380.
Patient problems: Insufficient Information
MAUDE report 24464277 ↗MalfunctionFeb 4, 2026Sparking
IT WAS REPORTED THAT THE FORCEPS WERE SPARKING AND SMOKING. INSULATION HAS SEPARATED FROM THE COLLAR WHERE IT GOES INTO THE HANDLE ON ALL ITEMS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257673 ↗MalfunctionFeb 4, 2026Sparking
IT WAS REPORTED THAT THE FORCEPS WERE SPARKING AND SMOKING. INSULATION HAS SEPARATED FROM THE COLLAR WHERE IT GOES INTO THE HANDLE ON ALL ITEMS .NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257636 ↗MalfunctionApr 29, 2025Smoking; Use of Device Problem; Smoking; Use of Device Problem
THE CUSTOMER REPORTED AN ISSUE WITH ONE OF THEIR BIPOLAR BURNERS PRIOR TO PROCEDURE. IT FUNCTIONED CORRECTLY WHEN TESTED ON A WET WIPE AND WAS PLACED IN A BAG READY FOR USE. HOWEVER, THE OPERATING ROOM NURSE NOTICED A BURNT SMELL AND REMOVED THE BURNER FROM THE BAG. UPON RETESTING, SMOKE EMERGED FROM THE END NEAR THE JAWS, AND THE BURNER SUBSEQUENTLY FAILED TO OPERATE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21931272 ↗MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Use of Device Problem; Electrical /Electronic Property Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY, ITEM SPARKED INSIDE OF PATIENT. THERE WAS NO ONE FROM KARL STORZ TO WITNESSED THIS. MANAGER, ONSITE ENDOSCOPY BELIEVES IT IS THE INSERT, AND WOULD LIKE ALL THREE ITEMS EVALUATED TOGETHER. FEMALE PATIENT NOT HARMED, CASE WAS NOT DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19912206 ↗InjuryAug 30, 2022Material Fragmentation; Material Fragmentation
INFORMATION RECIVED BY KARL STORZ INDICATES THAT THE DEVICE'S SHEATH WAS SEEN TO HAVE A SMALL SECTION MISSING ON THE TIP AND IT WAS NOT KNOWN IF IT WAS INSIDE THE PATIENT. THERE WAS NO WAY TO KNOW IF THE MISSING PIECE WAS MISSING BEFORE PROCEDURE.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 15324428 ↗InjuryNov 28, 2020Material Integrity Problem; Material Integrity Problem
ALLEGEDLY, THE PATIENT WAS UNDERGOING A HYSTERECTOMY PROCEDURE AND THE BURN WAS NOTICED BY DR. (B)(6) AT THE END OF THE PROCEDURE. THE INSTRUMENTS WERE NOT EXCHANGED OUT AND THE CASE WAS COMPLETED WITH NO TREATMENT NECESSARY.
Patient problems: Patient Problem/Medical Problem
MAUDE report 10915777 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page