Skip to content

ROBI 38610MLRecalls & reported problems

Karl Storz GmbH & Co. KG · RoBi® KELLY Forceps Insert · Class 2

All models of ROBI (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04048551328675), received between May 9, 2019 and Jun 30, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • No Display/Image2
  • Firing Problem1
  • Mechanical Problem1
  • Overheating of Device1
  • Sparking1
  • Temperature Problem1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 30, 2026No Display/Image

    IT WAS REPORTED THAT THE FORCEPS SCREW FELL INTO THE PATIENT. TO LOCATE THE SCREW, X-RAYS WERE TAKEN.

    Patient problems: Insufficient Information

    MAUDE report 25668590 ↗
  • InjuryJun 30, 2026No Display/Image

    IT WAS REPORTED THAT THE FORCEPS SCREW FELL INTO THE PATIENT. TO LOCATE THE SCREW, X-RAYS WERE TAKEN.

    Patient problems: Insufficient Information

    MAUDE report 25668094 ↗
  • MalfunctionJan 28, 2026Electrical /Electronic Property Problem; Temperature Problem

    IT WAS REPORTED THAT THE ITEM SPRAYS SPARKS DURING PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24202965 ↗
  • MalfunctionDec 16, 2025Sparking; Sparking

    IT WAS REPORTED: "FORCEP SPARKING AND SMOKING. INSULATION HAS SEPARATED FROM THE COLLAR WHERE IT GOES INTO THE HANDLE ON ALL ITEMS." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23817623 ↗
  • MalfunctionMay 14, 2025Thermal Decomposition of Device; Electrical /Electronic Property Problem; Overheating of Device; Thermal Decomposition of Device; Electrical /Electronic Property Problem; Overheating of Device

    IT WAS REPORTED THAT A ROBI CLAMP WAS BURNED DURING THE MEDICAL PROCEDURE. ACCORDING TO THE CUSTOMER, IT WAS USED IN A GYNECOLOGY PROCEDURE, WITH THE SAME BRAND OF AUTOCON HIGH FREQUENCY EQUIPMENT, USING THE RECOMMENDED ENERGY. THEY HAVE NOT IDENTIFIED WHAT COULD HAVE CAUSED THE FORCEPS TO BURN. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22020594 ↗
  • MalfunctionMay 9, 2019Mechanical Problem; Firing Problem; Mechanical Problem; Firing Problem

    ALLEGEDLY, THERE WAS A SPARK FROM THE TIP OF THE INSTRUMENT IN THE PATIENT DURING A PROCEDURE. IT WAS REPORTED THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8597855 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page