Skip to content

IMAGE 1 TH100Recalls & reported problems

Karl Storz GmbH & Co. KG · IMAGE1 S H3-Z · Class 2

Adverse event reports
20
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 04048551333136), received between Aug 13, 2021 and Aug 28, 2026 — classified as 2 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Erratic or Intermittent Display” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Erratic or Intermittent Display5
  • Display or Visual Feedback Problem3
  • Poor Quality Image3
  • No Apparent Adverse Event2
  • No Display/Image2
  • Optical Distortion2
  • Optical Problem2
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • Insufficient Information3

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2026Erratic or Intermittent Display

    Correction to the faulty product.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50499305 ↗
  • MalfunctionAug 25, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50439098 ↗
  • MalfunctionAug 6, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162790 ↗
  • MalfunctionJul 9, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25769179 ↗
  • MalfunctionApr 8, 2026Poor Quality Image; No Apparent Adverse Event

    IT WAS REPORTED THAT "THE CAMERA HEAD, SERIAL NUMBER (B)(6), BECAME FOGGED DURING THE FIRST PROCEDURE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24823060 ↗
  • MalfunctionMar 30, 2026Optical Distortion

    THE FOLLOWING WAS REPORTED: THE SCREEN IMAGING IS BLURRED; NORMAL FUNCTION IS RESTORED AFTER REPLACING THE CAMERA. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24719777 ↗
  • MalfunctionMar 6, 2026Poor Quality Image

    IT WAS REPORTED THAT "DURING TESTING WITH THE ENDOSCOPE CAMERA SYSTEM: AFTER A PERIOD OF USE, THE CAMERA IMAGE BECAME FOGGY, RESULTING IN UNCLEAR VIEW AND UNUSABLE CONDITION. THE CAMERA IMAGE WAS NORMAL, BUT THE APERTURE COULD NOT BE ADJUSTED, AFFECTING INTRAOPERATIVE OBSERVATION. THE DEVICE HAS BEEN TAKEN OUT OF SERVICE.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24537121 ↗
  • MalfunctionDec 30, 2025No Display/Image; No Apparent Adverse Event; No Display/Image; No Apparent Adverse Event

    IT WAS REPORTED THAT CAMERA WAS RECOGNIZED BY THE PROCESSOR, BUT THERE WAS NO IMAGE AND INTERFERENCE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23923568 ↗
  • InjuryDec 29, 2025Sparking; Sparking

    IT WAS REPORTED THAT THE SURGEON FELT A 'SPARK' OR A SHOCK WHEN USING THIS CAMERA AND THEY JUMPED. A STAFF MEMBER WITH THE SURGEON CLAIMS THERE WERE NO OTHER ELECTRICAL DEVICES, AND THE SURGEON WAS WEARING GLOVES. WE RECKON IT WAS SOMETHING ELSE OR THE SURGEON THEMSELVES THAT HAD THE ISSUE, AND NOT THE DEVICE. COULD BE STATIC. WE COMPLETED AN EST WITH APPLIED PARTS (ON THE CAMERA'S BARE METAL PART), AND IT PASSED". SINCE THE SURGEON FELT A 'SPARK' OR A SHOCK WHEN USING THIS CAMERA, THIS CASE IS REPORTABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23908240 ↗
  • MalfunctionDec 22, 2025Unexpected Shutdown; Unexpected Shutdown

    ACCORDING TO THE INFORMATION RECEIVED, THE CAMERA IS DEFECTIVE AND STOPS WORKING DURING OPERATION. IT WAS TESTED ON TWO DIFFERENT PROCESSORS. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23865648 ↗
  • MalfunctionDec 8, 2025Optical Distortion; Optical Distortion

    THE FOLLOWING WAS REPORTED: "CLOUDY IMAGE" NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23746951 ↗
  • MalfunctionOct 15, 2025Output Problem; Output Problem

    IT WAS REPORTED "BLACK SCREEN NO IMAGE DURING CASE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23302399 ↗
  • MalfunctionSep 18, 2025Signal Artifact/Noise; No Display/Image; Signal Artifact/Noise; No Display/Image

    IT WAS REPORTED THAT THE IMAGE FLICKERS OVER THE SCREEN AND LINES SOMETIMES APPEAR WHEN YOU WIGGLE THE CAMERA CABLE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23089753 ↗
  • MalfunctionAug 7, 2025Fracture; Optical Problem; Fracture; Optical Problem

    IT WAS REPORTED THAT THE "EQUIPMENT EXPERIENCES INTERFERENCE WHEN THE HEAD IS MOVED DURING SURGERY." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739010 ↗
  • MalfunctionMay 23, 2025Optical Problem; Optical Problem

    THE FOLLOWING WAS REPORTED: "BLURRY IMAGES." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22081473 ↗
  • MalfunctionMay 12, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE IMAGE SHUT OFF MID PROCEDURE (DATE, TYPE AND ADDITIONAL DETAILS NOT PROVIDED). NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005947 ↗
  • MalfunctionMar 5, 2025Erratic or Intermittent Display; Mechanical Problem; Poor Quality Image; Erratic or Intermittent Display; Mechanical Problem; Poor Quality Image

    IT WAS REPORTED THAT "DURING THE PROCEDURE, THE IMAGE JUMPS, AND IT IS NOT CLEAR. A DIFFERENT TH100 WAS USED TO COMPLETE THE PROCEDURE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21528677 ↗
  • MalfunctionFeb 10, 2023Image Display Error/Artifact; Image Display Error/Artifact

    THE EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT THE IMAGE BECOME BLACK AFTER 15 MINUTES. THE DEVICE NEEDS TO BE RESTARTED TO SOLVE THE ISSUE. AFTER RESTART, THERE IS A COLOR CASE, OR THE VIDEO PROCESSOR REPORTS AN INCOMPATIBLE CAMERA. NO HARM TO THE PATIENT, USER, OR THIRD REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16354038 ↗
  • MalfunctionJan 13, 2023Display or Visual Feedback Problem; Moisture or Humidity Problem; Display or Visual Feedback Problem; Moisture or Humidity Problem

    IT WAS REPORTED THAT THE IMAGE CUT OUT DURING CASE NO PATIENT INJURY. PROCEDURE WAS COMPLETED USING A BACK-UP CAMERA. THE SURGERY WAS DELAYED APPROXIMENTLY 5 TO 6 MINUTES. ADDITIONAL PATIENT INFORMATION IS NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16158999 ↗
  • InjuryAug 13, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem

    AS PER A MANUFACTURER INCIDENT REPORT WE RECEIVED FROM THE FACTORY IN (B)(4): DURING SURGERY, THE IMAGE OF THE CAMERA WAS INTERMITTENT. THE CUSTOMER SWITCHED TO OPEN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 12320991 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.