IMAGE1 CONNECT TC200USRecalls & reported problems
Karl Storz GmbH & Co. KG · CCU,MODULAR,IMAGE 1 CONNECT,US · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04048551356586), received between Feb 9, 2018 and Jul 30, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Erratic or Intermittent Display” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Erratic or Intermittent Display3
- Computer Software Problem1
- Device Operates Differently Than Expected1
- Display or Visual Feedback Problem1
- Programming Issue1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT "UNITS ARE GOING BLACK/BLANK WHEN A CAMERA IS PLUGGED IN. CUT OUT IN MIDDLE OF PROCEDURE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22666436 ↗MalfunctionMay 13, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT DURING A SINUS DIAGNOSTIC ENDOSCOPY PROCEDURE (CASE DATE UNKNOWN), IMAGE ON THE MONITOR AND SMARTSCREEN FLICKERED. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITHOUT ANY PATIENT IMPACT. CROSS REFERENCE MDR: 2027009-2025-00796.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22012524 ↗MalfunctionMay 13, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT DURING A SINUS DIAGNOSTIC ENDOSCOPY PROCEDURE (CASE DATE UNKNOWN), IMAGE ON THE MONITOR AND SMARTSCREEN FLICKERED. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITHOUT ANY PATIENT IMPACT. CROSS REFERENCE MDR: 2027009-2025-00792
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22012523 ↗MalfunctionMay 13, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT DURING A SINUS DIAGNOSTIC ENDOSCOPY PROCEDURE (CASE DATE UNKNOWN), IMAGE ON THE MONITOR AND SMARTSCREEN FLICKERED. THEY WERE ABLE TO COMPLETE THE PROCEDURE WITHOUT ANY PATIENT IMPACT. CROSS REFERENCE MDR: 2027009-2025-00794.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22012522 ↗MalfunctionFeb 9, 2018Device Operates Differently Than Expected
ALLEGEDLY, DURING SET UP FOR A LAPAROSCOPY PROCEDURE, THE LIGHT SOURCE, WITH A LIGHT CABLE CONNECTED, WAS IN STAND-BY MODE AND THE LIGHT CABLE WAS NOT CONNECTED TO A SCOPE. WHEN THE CAMERA HEAD WAS CONNECTED TO THE TS200US CAMERA CONTROL UNIT (CCU) , THE LIGHT SOURCE AUTOMATICALLY WENT TO PREVIOUS SETTING OF 100% LIGHT INTENSITY AND, BECAUSE THE LIGHT CABLE WAS NOT CONNECTED TO A SCOPE AND IT WAS IN CONTACT WITH THE PATIENT'S DRAPE, THE DRAPE WAS BURNED. THERE WAS NO IMPACT ON PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7261844 ↗MalfunctionFeb 9, 2018Computer Software Problem; Programming Issue; Computer Software Problem; Programming Issue
ALLEGEDLY, DURING SET UP FOR A HYSTEROSCOPY PROCEDURE, THE LIGHT SOURCE, WITH A LIGHT CABLE CONNECTED, WAS IN STAND-BY MODE AND THE LIGHT CABLE WAS NOT CONNECTED TO A SCOPE. WHEN THE CAMERA HEAD WAS CONNECTED TO THE TS200US CAMERA CONTROL UNIT (CCU) , THE LIGHT SOURCE AUTOMATICALLY WENT TO PREVIOUS SETTING OF 100% LIGHT INTENSITY AND, BECAUSE THE LIGHT CABLE WAS NOT CONNECTED TO A SCOPE AND IT WAS IN CONTACT WITH THE PATIENT'S DRAPE, THE DRAPE WAS BURNED. THERE WAS NO IMPACT ON PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7261842 ↗
FDA clearance and classification
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the IMAGE1 CONNECT TC200US
Has the IMAGE1 CONNECT TC200US been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551356586), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the IMAGE1 CONNECT TC200US?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 9, 2018 and Jul 30, 2025. In those reports, the most common specific device problems are “Erratic or Intermittent Display” (3), “Computer Software Problem” (1) and “Device Operates Differently Than Expected” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the IMAGE1 CONNECT TC200US?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the IMAGE1 CONNECT TC200US, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand IMAGE1 CONNECT, model TC200US). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,392 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page