IMAGE 1 TH111Recalls & reported problems
Karl Storz GmbH & Co. KG · IMAGE1 HX-P · Class 2
All models of IMAGE 1 (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04048551358931), received between Oct 6, 2023 and May 22, 2026 — classified as 4 injuries, 6 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image7
- Erratic or Intermittent Display4
- Display or Visual Feedback Problem1
- Image Display Error/Artifact1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2026No Display/Image
IT WAS REPORTED THAT "WHEN THE CAMERA HEAD ARE NOT MOVED, THE DISPLAY DOES NOT APPEAR ON THE MONITOR.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25262712 ↗MalfunctionFeb 10, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT A COMPLETE SIGNAL LOSS (BLUE SCREEN) DURING AN ONGOING COELIOSCOPY PROCEDURE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24306310 ↗MalfunctionDec 19, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT "THE SCREEN SUDDENLY DEFAULTED TO THE HOME SCREEN AS IF THE CAMERA HE'D HAD BEEN REMOVED. THIS HAPPENED SEVERAL TIMES DURING THE PROCEDURE, HOWEVER THE IMAGE WAS QUICKLY RESTORED THEREAFTER.". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23858114 ↗MalfunctionDec 10, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE ENDOSOPIC IMAGE WAS LOST DURING USE. STAFF WERE THEN REQUIRED TO RESOLVE THIS ISSUE BY PLUGGING ANOTHER UROLOGY PENDULUM CAMERA HEAD INTO THE STACK IN USE, WHICH THEN ALLOWED THE SURGERY TO PROCEED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770495 ↗MalfunctionDec 10, 2025Erratic or Intermittent Display; Signal Artifact/Noise; Erratic or Intermittent Display
IT WAS REPORTED THAT THE CAMERA HEAD TH111 WAS PURCHASED FOR A UROLOGY STACK. WHEN THE STACK IS USED WITH LASERS, CLINICAL STAFF HAVE REPORTED INTERFERENCE ON THE MONITOR. THE SCREEN WAS FLICKERING AND HAD ARTEFACTS. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770464 ↗MalfunctionMay 12, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE DEVICE DID NOT WORK WHEN THE CUSTOMER TRIED TO USE IT FOR THE FIRST TIME. THEY DOUBLE CHECKED IT AGAIN, AND STILL NO IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22004084 ↗InjuryMay 2, 2025Erratic or Intermittent Display; No Display/Image; Image Display Error/Artifact; Erratic or Intermittent Display; No Display/Image; Image Display Error/Artifact
IT WAS REPORTED THAT THE IMAGE IS DISTORTED ON SCREEN WHICH MAKES THE PICTURE DIFFICULT TO SEE. IT WAS FURTHERMORE STATED THAT THIS WAS NOTICE DURING A PATIENT PROCEDURE, BUT THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 25-APR-2025. REGARDING THE REPORTED ERROR IT WAS STATED "NO IMAGE OUTPUT AND SCREEN DISPLAYED WITH MULTIPLE LINES FLICKERING WHEN THE CAMERA WAS USED ON THE PATIENT". FURTHERMORE, IT WAS CONFIRMED AGAIN THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT GOT DISCHARGED HOME SAFE, BUT IT WAS ALSO INDICATED THAT THE PROCEDURE WAS PROLONGED BETWEEN 30 AND 60 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21952188 ↗InjuryMay 2, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE IMAGE IS DISTORTED ON SCREEN WHICH MAKES THE PICTURE DIFFICULT TO SEE. IT WAS FURTHERMORE STATED THAT THIS WAS NOTICE DURING A PATIENT PROCEDURE, BUT THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 25-APR-2025. REGARDING THE REPORTED ERROR IT WAS STATED "NO IMAGE OUTPUT AND SCREEN DISPLAYED WITH MULTIPLE LINES FLICKERING WHEN THE CAMERA WAS USED ON THE PATIENT". FURTHERMORE, IT WAS CONFIRMED AGAIN THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT GOT DISCHARGED HOME SAFE, BUT IT WAS ALSO INDICATED THAT THE PROCEDURE WAS PROLONGED BETWEEN 30 AND 60 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21952186 ↗InjuryOct 6, 2023Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
IT WAS REPORTED THAT "MID SURGERY, WHEN USING THE CAMERA SYSTEM ALONGSIDE A DIATHERMY, THE STACK LOST THE PICTURE ON THE LARGE MONITOR DUE TO INTERFERENCE". FURTHERMORE: "ELONGATED AND INCOMPLETE PROCEDURE. POSSIBLE REQUIREMENT OF SECONDARY PROCEDURE."
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17886677 ↗InjuryOct 6, 2023Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT "MID SURGERY, WHEN USING THE CAMERA SYSTEM ALONGSIDE A DIATHERMY, THE STACK LOST THE PICTURE ON THE LARGE MONITOR DUE TO INTERFERENCE". FURTHERMORE: "ELONGATED AND INCOMPLETE PROCEDURE. POSSIBLE REQUIREMENT OF SECONDARY PROCEDURE."
Patient problems: Insufficient Information
MAUDE report 17886673 ↗
FDA clearance and classification
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the IMAGE 1 TH111
Has the IMAGE 1 TH111 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551358931), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the IMAGE 1 TH111?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Oct 6, 2023 and May 22, 2026. In those reports, the most common specific device problems are “No Display/Image” (7), “Erratic or Intermittent Display” (4) and “Display or Visual Feedback Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the IMAGE 1 TH111?
None of the reports are classified as deaths, 4 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the IMAGE 1 TH111, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand IMAGE 1, model TH111). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,392 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page