IMAGE 1 S TH120Recalls & reported problems

Karl Storz GmbH & Co. KG · IMAGE1 S 4U · Class 2

All models of IMAGE 1 S (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Karl Storz IMAGE1 S TH120 endoscopic video imaging system is used in gastroenterology and urology procedures. Most of its 30 reports describe picture problems during surgery — blank, flickering, blurry or dark images, and times when the camera head or its cable was thought to have come unplugged, in one case leading surgeons to convert a laparoscopic procedure to an open one. Three reports describe injuries, and none describe deaths. No recall names this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04048551370506), received between Mar 9, 2021 and Jun 19, 2026 — classified as 3 injuries, 22 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 4 reportsAug 2025: 2 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image9
  • Display or Visual Feedback Problem4
  • Adverse Event Without Identified Device or Use Problem3
  • Erratic or Intermittent Display3
  • Insufficient Device Problem Information3
  • No Apparent Adverse Event3
  • Optical Problem3
  • Defective Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information2
  • Superficial (First Degree) Burn2

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2026No Display/Image

    THE FOLLOWING WAS REPORTED: "CAMERA HEAD WAS FAULTY, OR CAMERA UNPLUGGED. THE PROCEDURE WAS CONVERTED FROM LAP TO OPEN. THE CASE WAS COMPLETED. UPON FURTHER INVESTIGATION IT IS UNCLEAR WHETHER THE CAMERA HEAD WAS FAULTY, OR IF IT WAS JUST THE CAMERA BOX THAT WAS UNPLUGGED." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25564188 ↗
  • MalfunctionApr 28, 2026No Display/Image

    IT WAS REPORTED THAT INTERMITTENT IMAGE PROBLEMS OCCURED DURING SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25020359 ↗
  • MalfunctionApr 2, 2026No Display/Image

    ACCORDING TO THE INFORMATION RECEIVED, NO IMAGE WAS OBSERVED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24771927 ↗
  • MalfunctionJan 27, 2026Optical Problem

    IT WAS REPORTED THAT THE IMAGE DISPLAYS BLURRY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24193748 ↗
  • MalfunctionJan 16, 2026Display or Visual Feedback Problem

    THE FOLLOWING WAS REPORTED: "CHECKED AND FOUND NO IMAGE DISPLAY WHEN CONNECTED TO CCU CONNECTOR CONTACT POINTS." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24104876 ↗
  • MalfunctionNov 26, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE DEVICE HAS A BLURRY IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23660628 ↗
  • MalfunctionNov 12, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT, PRIOR TO THE PROCEDURE, THE IMAGE FLICKERED WHEN THE CAMERA HEAD WAS CONNECTED TO THE CAMERA BOX. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23538503 ↗
  • InjurySep 18, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT "WHEN THE CAMERA HEAD IS CONNECTED TO THE CAMERA BOX, THE IMAGE FLICKERS, AND THE SURGEON CANNOT SEE THE PATIENT'S ANATOMY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. THE PATIENT WAS MOVED ANESTHETIZED, TO ANOTHER OR FOR THE PROCEDURE TO BE COMPLETED. THEY PULLED A NEW CAMERA HEAD FOR THE CASE TO MOVE ON IN A DIFFERENT OR. THERE WAS A DELAY OF AT LEAST ONE HOUR. CROSS REFERENCE MDR: 2027009-2025-01622.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23087257 ↗
  • InjurySep 18, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT "WHEN THE CAMERA HEAD IS CONNECTED TO THE CAMERA BOX, THE IMAGE FLICKERS, AND THE SURGEON CANNOT SEE THE PATIENT'S ANATOMY". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. THE PATIENT WAS MOVED ANESTHETIZED, TO ANOTHER OR FOR THE PROCEDURE TO BE COMPLETED. THEY PULLED A NEW CAMERA HEAD FOR THE CASE TO MOVE ON IN A DIFFERENT OR. THERE WAS A DELAY OF AT LEAST ONE HOUR. CROSS REFERENCE MDR: 2027009-2025-01623.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23087134 ↗
  • MalfunctionAug 21, 2025Optical Problem; Optical Problem

    THE FOLLOWING WAS REPORTED: CAMERA WILL NOT PRODUCE ANY IMAGE WHEN CONNECTED TO 4U LINK, REMAINS ON MAIN MENU SCREEN. DOCTOR HAD TO SWITCH CAMERAS AT THE BEGINNING OF CYSTO CASE, CAUSED NO HARM JUST A DELAY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22864939 ↗
  • MalfunctionAug 15, 2025No Display/Image; No Apparent Adverse Event; No Display/Image; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CAMERA HEAD CABLE HAD A LOOSE CONNECTION, WHICH CAUSED THE SCREEN TO REPEATEDLY TURN BLACK. THE SURGICAL PROLONGATION TOOK LESS THAN 15 MINUTES. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22819822 ↗
  • MalfunctionJul 24, 2025Display or Visual Feedback Problem; No Apparent Adverse Event; Display or Visual Feedback Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CAMERA PRESENTED A SHAKING IMAGE AND SOMETIME A BLACK SCREEN. THE SHAKING IS NOT PERMANENT. IT OCCURED FEW TIMES THE SOUND OF THE UNITS ARE VERY ACUTE AND IS SUGGESTING A FAN PROBLEM. THIS SITUATION OCCURED DURING USE AND DURING A TEST IN OUR TECHNICAL DEPARTMENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY OR PROLONGATION OF SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22589469 ↗
  • MalfunctionJul 24, 2025Display or Visual Feedback Problem; No Apparent Adverse Event; Display or Visual Feedback Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CAMERA PRESENTED A SHAKING IMAGE AND SOMETIME A BLACK SCREEN. THE SHAKING IS NOT PERMANENT. IT OCCURED FEW TIMES THE SOUND OF THE UNITS ARE VERY ACUTE AND IS SUGGESTING A FAN PROBLEM. THIS SITUATION OCCURED DURING USE AND DURING A TEST IN OUR TECHNICAL DEPARTMENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY OR PROLONGATION OF SURGERY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22589313 ↗
  • MalfunctionJul 21, 2025No Display/Image; Material Split, Cut or Torn; No Display/Image; Material Split, Cut or Torn

    ACCORDING TO THE INFORMATION RECEIVED, AFTER ONLY 4 USES (AFTER RETURNED FROM SERVICE) THE SAME PROBLEM OCCURRED. THE CAMERA DOES NOT GIVE IMAGE WHEN CONNECTED. NO PATIENT INVOLVEMENT REPORTED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22558065 ↗
  • MalfunctionJul 10, 2025No Display/Image; No Display/Image

    IT WAS REPORTED THAT INTERMITTENT IMAGES OCCURRED DURING THE PROCEDURE, BUT THEY WERE ABLE TO COMPLETE THE PROCEDURE BY USING A BACKUP CAMERA HEAD. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22472959 ↗
  • MalfunctionJun 24, 2025Optical Problem; Optical Problem

    ACCORDING TO THE INFORMATION RECEIVED, NOT GIVING ANY PICTURE AFTER SHORT TIME. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22323685 ↗
  • MalfunctionMay 31, 2025Output Problem; Output Problem

    IT WAS REPORTED THAT DURING THE USE, THE DEVICE TURNS OFF. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22120745 ↗
  • MalfunctionMay 2, 2025No Display/Image; Defective Component; No Display/Image; Defective Component

    IT WAS REPORTED THAT THERE IS NO IMAGE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21954442 ↗
  • MalfunctionMay 2, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT DURING A VAAFT PROCEDURE, THE IMAGE ON THE MONITOR WAS LOST FOR A BRIEF SECOND, TWICE IN SHORT SUCCESSION. THE SURGEON SAID IT WAS WHILE USING DIATHERMY. [[?]] PICTURE FLICKERED WHEN DIATHERMY MACHINE WAS BEING USED THEN WENT OFF WHILE USING DIATHERMY. FURTHERMORE, IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT, USER OR THIRD PARTY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21952233 ↗
  • MalfunctionMay 1, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT DURING A PROCEDURE (TYPE NOT PROVIDED), THE IMAGE WENT BLACK AND THE WHITE BALANCE WAS INCOMPLETE. NO HARM TO PATIENT, USER OR THIRD OCCURRED. THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE WITH A DELAY OF 10 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942953 ↗
  • MalfunctionApr 28, 2025No Display/Image; Defective Component; No Display/Image; Defective Component

    IT WAS REPORTED THAT THE "CAMERA HEAD WITH INTERMITTENT FAULT NOT DISPLAYING IMAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21923281 ↗
  • MalfunctionFeb 6, 2025No Display/Image; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; No Display/Image; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT LOSS OF IMAGE ON SURGICAL MONITOR 2 LOTS OR1AIR HAPPENED AT THE SAME TIME IN THE SAME THEATRE ON THE SAME PATIENT. THERE WAS A TEMPORARY BLANK SCREEN FOR APPROX. 15 SECONDS WHILST SURGERY WAS IN PROGRESS THERE WAS A FAULTY MONITOR - FURTHER INVESTIGATION NEEDED AS THIS MAY ALSO BE AS A RESULT OF AN ISSUE WITH THE TC201EN, OR TH120, OR OR1AIR (WD300 / AVM). NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 9610617-2025-00147. 9610617-2025-00149.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21329695 ↗
  • MalfunctionNov 22, 2024Break; Intermittent Communication Failure; Break; Intermittent Communication Failure

    IT WAS REPORTED THAT DURING THE PROCEDURE THE PICTURE OF THE CAMERA IS GONE. AFTER A SHORT PERIOD THE PICTURE IS BACK. NO HARM TO PATIENT, USER OR THIRD REPORTED. NO NEGATIVE CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20763152 ↗
  • MalfunctionMar 9, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    PER CONVERSATION WITH KARL STORZ REPRESENTATIVE AND THE FACILITY, AFTER A SEPTORHINOPLASTY PROCEDURE A "SMALL REDDISH BLEMISH" WAS NOTED POST OP WITH NO AWARENESS TO A PATIENT FOLLOW UP.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 11449083 ↗
  • MalfunctionMar 9, 2021Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    PER CONVERSATION WITH KARL STORZ REPRESENTATIVE AND THE FACILITY, AFTER A SEPTORHINOPLASTY PROCEDURE A "SMALL REDDISH BLEMISH" WAS NOTED POST OP WITH NO AWARENESS TO A PATIENT FOLLOW UP.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 11449076 ↗

FDA clearance and classification

The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the IMAGE 1 S TH120

Has the IMAGE 1 S TH120 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04048551370506), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the IMAGE 1 S TH120?

The FDA's MAUDE database lists 25 adverse event reports for this device, received between Mar 9, 2021 and Jun 19, 2026. In those reports, the most common specific device problems are “No Display/Image” (9), “Display or Visual Feedback Problem” (4) and “Erratic or Intermittent Display” (3); the most common patient problem is “Superficial (First Degree) Burn” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the IMAGE 1 S TH120?

None of the reports are classified as deaths, 3 as injuries and 22 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the IMAGE 1 S TH120, and is it still sold?

The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand IMAGE 1 S, model TH120). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.