TELE PACK TP101Recalls & reported problems
Karl Storz GmbH & Co. KG · TELE PACK + · Class 2
All models of TELE PACK (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Karl Storz Tele Pack is a video imaging unit used with endoscopes in gastroenterology and urology. The 42 reports that mention its identifier mostly describe display problems such as blank or intermittent images and error messages, with
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 04048551425701), received between Feb 19, 2025 and Jul 7, 2026 — classified as 14 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem6
- Erratic or Intermittent Display5
- No Apparent Adverse Event5
- Data Problem4
- No Display/Image4
- Defective Component3
- Defibrillation/Stimulation Problem2
- Detachment of Device or Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- Insufficient Information6
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2026Erratic or Intermittent Display
IT WAS REPORTED: "INTERMITTENT IMAGE IN THE C-PORT, NO IMAGES ON SLIMLINE AND SINGLE USE SCOPES. ALL SCOPES HAVE IMAGES ON THE OTHER TELEPAK". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25736521 ↗MalfunctionJul 7, 2026Loose or Intermittent Connection
THE FOLLOWING WAS REPORTED: "DISPLAYING AN INTERNAL ERROR MESSAGE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25736513 ↗MalfunctionJun 16, 2026Erratic or Intermittent Display
IT WAS REPORTED "CAMERA SOCKET ISSUES - IMAGES CUT OUT INTERMITTENTLY, INTERMITTENT CAMERA HEAD NOT RECOGNISED ERROR MESSAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25521741 ↗InjuryMay 12, 2026Defibrillation/Stimulation Problem
IT WAS REPORTED THAT THE STROBOSCOPY OUTPUT IS NOT FUNCTIONING. WHEN THE MIC IS CONNECTED AND THE FOOTPEDAL IS DEPRESSED, NONE OF THE ON-SCREEN READOUTS ARE PORTRAYED. IT REMAINS BLANK AND APPEARS AS IF THE MIC AND LIGHT ARE NOT TRACKING THE MUCOSAL WAVE. I HAVE TRIED OTHER MICS AND FOOTPEDALS ON THAT UNIT AND THEY DIDN'T WORK EITHER AND I ALSO TRIED THEIR MIC AND FOOTPEDAL ON ANOTHER UNIT AND IT WORKS FINE. IT WAS NOT POSSIBLE TO COMPLETE THE STROBOSCOPY PORTION OF THE PROCEDURE EFFECTIVELY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25155679 ↗InjuryMay 12, 2026Defibrillation/Stimulation Problem
IT WAS REPORTED THAT THE STROBOSCOPY OUTPUT IS NOT FUNCTIONING. WHEN THE MIC IS CONNECTED AND THE FOOTPEDAL IS DEPRESSED, NONE OF THE ON-SCREEN READOUTS ARE PORTRAYED. IT REMAINS BLANK AND APPEARS AS IF THE MIC AND LIGHT ARE NOT TRACKING THE MUCOSAL WAVE. I HAVE TRIED OTHER MICS AND FOOTPEDALS ON THAT UNIT AND THEY DIDN'T WORK EITHER AND I ALSO TRIED THEIR MIC AND FOOTPEDAL ON ANOTHER UNIT AND IT WORKS FINE. IT WAS NOT POSSIBLE TO COMPLETE THE STROBOSCOPY PORTION OF THE PROCEDURE EFFECTIVELY. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25155677 ↗MalfunctionMar 13, 2026Output Problem
THE FOLLOWING WAS REPORTED: "PROVIDING ERROR MESSAGE UPPER RIGHT CORNER - SYSTEM NOT RESPONDING PLEASE CYCLE IF ISSUE PERSIST." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591110 ↗InjuryMar 6, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT "GETTING MENU TEMPORARILY UNAVAILABLE MESSAGE CONTINUOUSLY, ETHERNET PORT DAMAGED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, THEY WERE NOT ABLE TO FINISH THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24539077 ↗InjuryMar 6, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT "GETTING MENU TEMPORARILY UNAVAILABLE MESSAGE CONTINUOUSLY, ETHERNET PORT DAMAGED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, THEY WERE NOT ABLE TO FINISH THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24538918 ↗InjuryMar 6, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT "GETTING MENU TEMPORARILY UNAVAILABLE MESSAGE CONTINUOUSLY, ETHERNET PORT DAMAGED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, THEY WERE NOT ABLE TO FINISH THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24538904 ↗InjuryMar 6, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT "GETTING MENU TEMPORARILY UNAVAILABLE MESSAGE CONTINUOUSLY, ETHERNET PORT DAMAGED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, THEY WERE NOT ABLE TO FINISH THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24538827 ↗InjuryMar 6, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT "GETTING MENU TEMPORARILY UNAVAILABLE MESSAGE CONTINUOUSLY, ETHERNET PORT DAMAGED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. HOWEVER, THEY WERE NOT ABLE TO FINISH THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24538747 ↗MalfunctionFeb 25, 2026Optical Distortion
IT WAS REPORTED THAT FAULTS: DISPLAY GOES BLANK INTERMITTENTLY DURING PROCEDURE, THE PROCEDURE WAS EXTENDED LESS THAN 30 MINUTES AND THE HOSPITAL STAY WAS NOT EXTENDED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24449392 ↗MalfunctionFeb 24, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT "FAULTS: DISPLAY GOES BLANK INTERMITTENTLY DURING PROCEDURE, THE PROCEDURE WAS EXTENDED LESS THAN 30 MINUTES AND THE HOSPITAL STAY WAS NOT EXTENDED". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24436873 ↗MalfunctionFeb 12, 2026Erratic or Intermittent Display
IT WAS REPORTED THAT "THE IMAGE ON THE SCREEN IS FLASHING AND CUTS OUT". ACCORDING TO THE INFORMATION RECEIVED, THIS HAPPENED DURING A VETERINARY PROCEDURE AND THE PROCEDURE COULD BE COMPLETED WITH A PROLONGATION OF LESS THAN 15 MINUTES. THERE WAS NO INJURY OR IMPACT TO ANY HUMAN REPORTED. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. THE PRODUCT CONCERNED (TP101) IS ALSO INTENDED FOR USE DURING HUMAN MEDICAL PROCEDURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24321245 ↗MalfunctionJan 30, 2026No Display/Image
IT WAS REPORTED: "NO IMAGE / ERROR MESSAGE". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24226186 ↗InjuryJan 21, 2026Detachment of Device or Device Component
THE FOLLOWING WAS REPORTED: "ONE SIDE OF THE OUTER SHEATH'S LOCKING PLATE DISLODGED INTO THE BODY. IT WAS IMMEDIATELY RETRIEVED, BUT THE OTHER SIDE WAS FOUND MISSING. NO SHADOW WAS VISIBLE ON X-RAY, LEADING TO A DETERMINATION THAT IT WAS NOT PRESENT. IT IS HIGHLY LIKELY THAT IT HAD BEEN DAMAGED PREVIOUSLY."
Patient problems: Insufficient Information
MAUDE report 24141598 ↗InjuryJan 21, 2026Detachment of Device or Device Component
THE FOLLOWING WAS REPORTED: "ONE SIDE OF THE OUTER SHEATH'S LOCKING PLATE DISLODGED INTO THE BODY. IT WAS IMMEDIATELY RETRIEVED, BUT THE OTHER SIDE WAS FOUND MISSING. NO SHADOW WAS VISIBLE ON X-RAY, LEADING TO A DETERMINATION THAT IT WAS NOT PRESENT. IT IS HIGHLY LIKELY THAT IT HAD BEEN DAMAGED PREVIOUSLY."
Patient problems: Insufficient Information
MAUDE report 24141597 ↗MalfunctionJan 21, 2026Erratic or Intermittent Display; No Apparent Adverse Event
ACCORDING TO THE INFORMATION RECEIVED, INTERMITTENT SIGNAL INTERRUPTIONS OCCURRING. THE SIGNAL INTERRUPTIONS OCCURRED DURING A PRESENTATION OF THE DEVICE WITH THE STAFF AND NOT DURING A MEDICAL PROCEDURE, THERE WAS NO PATIENT INVOLVED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED. CROSS REFERENCE MDR: 2027009-2026-00095.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24141589 ↗MalfunctionDec 22, 2025No Display/Image; Defective Component; No Display/Image; Defective Component
THE EVENT OCCURRED IN SINGAPORE. IT WAS REPORTED THAT "THE UNIT HAS NO DISPLAY WHEN POWERED UP, NEED TO BE REBOOTED BEFORE IMAGE IS PRESENT". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23873987 ↗MalfunctionDec 20, 2025Erratic or Intermittent Display; No Apparent Adverse Event; Erratic or Intermittent Display; No Apparent Adverse Event
ACCORDING TO THE INFORMATION RECEIVED, WHEN TESTED THE NEW 11101HDK SCOPE THAT ARRIVED WITH THE TP101, THE SCREEN WAS FLICKERING AND GIVING ME A GRAINY IMAGE. THE SCOPE IS BRAND NEW AND I TESTED IT ON ANOTHER TP101 THAT THE CUSTOMER HAD THERE IN OFFICE AND THE IMAGE WAS PERFECT. NO PATIENT INVOLVEMENT. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23858426 ↗MalfunctionDec 20, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT "THEY ARE NOT GETTING AN IMAGE HAPPENING ON AND OFF IN THE BEGINNING NOW IS HAPPENING ALL THE TIME, AND THEY DO NOT HAVE AN IMAGE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23858425 ↗MalfunctionDec 8, 2025Failure to Power Up; Failure to Power Up
THE FOLLOWING WAS REPORTED: "NO IMAGE WHEN PLUGGED IN." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23747081 ↗InjuryNov 26, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE FOLLOWING WAS REPORTED: "IMAGE WENT OUT, TO A BLUE SCREEN MULTIPLE TIMES DURING A PROCEDURE. THEY BOOTED IT SEVERAL TIMES, SEEMS TO WORK INTERMITTENTLY CUSTOMER IS COMPLAINING THAT IT TAKES LONGER THAN NORMAL, ALMOST TOO LONG FOR THE SCOPE AND CAMERA HEAD TO PRODUCE AN IMAGE." NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED. DUE TO THE MALFUNCTION OF THE DEVICE, THE SURGERY HAD TO BE ABANDONED. A BACKUP SYSTEM WAS NOT AVAILABLE. THEREFORE, THIS CASE IS NOW CLASSIFIED AS AN ADVERSE EVENT AND STILL REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23662958 ↗MalfunctionNov 3, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT "THE ADULT SCOPE PORT ISN'T READING, SO THERE'S NO IMAGE WHEN THEY PLUG IN THE ADULT SCOPE IN. BUT OTHER ONES DO GET IMAGES. THEY'VE TRIED MULTIPLE SCOPES AND IT'S DOING THE SAME THING". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452295 ↗MalfunctionOct 29, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ACCORDING TO THE INFORMATION RECEIVED, STOPPED WORKING INCOMPATIBLE HEAD NO IMAGE BLUE SCREEN, LIGHT ON SCOPE IS FLICKERING. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23416795 ↗
FDA clearance and classification
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TELE PACK TP101
Has the TELE PACK TP101 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04048551425701), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TELE PACK TP101?
The FDA's MAUDE database lists 34 adverse event reports for this device, received between Feb 19, 2025 and Jul 7, 2026. In those reports, the most common specific device problems are “Electrical /Electronic Property Problem” (6), “Erratic or Intermittent Display” (5) and “Data Problem” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TELE PACK TP101?
None of the reports are classified as deaths, 14 as injuries and 20 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TELE PACK TP101, and is it still sold?
The FDA's device database lists Karl Storz GmbH & Co. KG as the company responsible for it (brand TELE PACK, model TP101). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,392 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page