ERBE 20193-072Recalls & reported problems
Erbe Elektromedizin GmbH · ERBE MONOPlate 140 single; 136cm² · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04050147005143), received between Jul 12, 2022 and Jul 12, 2022 — classified as 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- No Apparent Adverse Event1
Patient problems
- Burn(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 12, 2022No Apparent Adverse Event; No Apparent Adverse Event
ON JUNE 21ST, 2022, WE HAVE BEEN INFORMED ABOUT AN INCIDENT INVOLVING A DISPERSIVE ELECTRODE AT UNIVERSITY HOSPITAL (B)(6). A MONITORING DISPERSIVE ELECTRODE ERBE NESSY PLATE 170 (MODEL XL31A/1) AND AN ABLATION DEVICE HAT 500 GENERATOR WERE USED. THE PROCEDURE LASTED FOR 164 MINUTES. THE INITIAL REPORTER STATED THAT [TRANSLATED FROM GERMAN LANGUAGE TO ENGLISH LANGUAGE] "DURING ELECTROPHYSIOLOGICAL PROCEDURE WITH ABLATION TREATMENT USING RADIOFREQUENCY CURRENT, UNEXPECTED BURNS OCCURRED IN THE AREA OF THE NEUTRAL ELECTRODE. AFTER REMOVAL, THE BURNS ON THE EDGE OF THE ELECTRODE BECAME VISIBLE AND WERE TREATED IMMEDIATELY, AND A CORRESPONDING CONSULTATION REQUEST WAS MADE TO THE COLLEAGUES IN TRAUMA SURGERY. IT WAS AN UNCOMPLICATED REGULAR PROCEDURE WITHOUT SIGNIFICANT ABNORMALITIES. EVEN IN THE FOLLOW-UP INSPECTION, NO ACCUMULATIONS OF LIQUID, INCORRECTLY CONNECTED CABLES OR OBVIOUS DEFECTS IN THE DEVICES COULD BE FOUND. THE AFFECTED NEUTRAL ELECTRODE (INCLUDING CABLE) WAS SECURED AND KEPT SAFE. NO OBVIOUS MATERIAL DEFECT COULD BE IDENTIFIED. THE ABLATION INCLUDING SCLEROTIZATION OF THE ADDITIONAL PATHWAY WAS SUCCESSFUL. THE FOLLOWING PRODUCTS HAVE BEEN USED FOR THE PROCEDURE: ABLATI
Patient problems: Burn(s)
MAUDE report 14984099 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page