Skip to content

NA 815033Recalls & reported problems

Richard Wolf GmbH · HF CONNECTION CABLE MONO L 3M for plug Bovie/Valleylab/Erbe Int., reusable · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04055207000236), received between Nov 16, 2018 and Feb 9, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Electrical Shorting1
  • Material Separation1
  • Sparking1
  • Temperature Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2024Material Separation; Electrical Shorting; Material Separation; Electrical Shorting

    AN IMPORTER HAS INFORMED RICHARD WOLF GMBH (RWGMBH) OF AN ISSUE REGARDING A HF CONNECTION CABLE MONO L 3M, PART ID: 815.033, SN/BATCH# UNKNOWN. THE INFORMATION WAS RECEIVED FROM A "USER-NOTIFICATION" REPORT NO. (B)(4). ACCORDING TO THE REPORT, THE CORD SHORTED OUT AND BLEW APART. NO FURTHER INFORMATION WAS PROVIDED REGARDING THE PATIENT'S OUTCOME OR ANY OTHER RISKS OR INJURIES. IT IS ALSO NOT KNOWN IF THERE WAS A TIME DELAY IN THE PROCEDURE.

    Patient problems: Insufficient Information

    MAUDE report 18678857 ↗
  • MalfunctionOct 29, 2020Break; Temperature Problem; Break; Temperature Problem

    ON FEBRUARY 4TH, 2021, RWMIC RECEIVED THE FOLLOWING INFORMATION FROM THE SALES REP: IN DISCUSSIONS WITH THE USER FACILITY AND THE DOCTOR IT WAS DETERMINED THAT IT WAS A MEDTRONIC GENERATOR CAUSING THE ISSUE. THE CUSTOMER WAS HAVING TO USE 3-4 LOOPS PER TURP. WE EXCHANGED OUT THE GENERATOR IN THE CYSTO ROOM AND FOUND IT WAS THE MEDTRONIC GENERATOR THAT WAS CAUSING THE LOOPS TO BURN OUT. THE CUSTOMER HAS SEND THE MEDTRONIC GENERATOR FOR REPAIR AND IS USING ONE OUT OF ANOTHER ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10756256 ↗
  • MalfunctionNov 16, 2018Sparking; Sparking

    FOLLOW-UP REPORT #2 IS TO PROVIDE FDA WITH NEW AND CHANGED INFORMATION. USER FACILITY RETURNED THE DEVICE TO RWMIC ON (B)(6) 2018, AND THE EVALUATION WAS COMPLETED ON (B)(6) 2019. THE CABLE WAS INSPECTED AND THE CABLE WAS FOUND TO BE BROKEN. THE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE WEAR AND TEAR. THE CAUSE FOR THE WIRE BREAKAGE IS THE USE OF THE CABLE. UP TO THE ABSOLUTE WEAR/ FAILURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8080385 ↗

FDA clearance and classification

The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the NA 815033

Has the NA 815033 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04055207000236), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 815033?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Nov 16, 2018 and Feb 9, 2024. In those reports, the most common specific device problems are “Break” (1), “Electrical Shorting” (1) and “Material Separation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 815033?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the NA 815033, and is it still sold?

The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand NA, model 815033). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page