Skip to content

NA 82306Recalls & reported problems

Richard Wolf GmbH · COAG ELECTRODE MONO 6FR WL 400MM for cystoscopy, electrode shape: button, PACK=1 PC, reusable · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207008102), received between Nov 16, 2018 and Nov 16, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Twisted/Bent1
  • Sparking1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 16, 2018Sparking; Material Twisted/Bent; Sparking; Material Twisted/Bent

    USER FACILITY RETURNED THE DEVICE TO RWMIC ON OCTOBER 29, 2018, AND THE EVALUATION WAS COMPLETED ON DECEMBER 04, 2018. THE DEVICE APPEARED TO BE USED - ELECTRODE WAS SEVERELY BENT AND DID NOT FIT FOR USE. THE REPORTED CONDITION WAS VERIFIED USING VISUAL, FUNCTIONAL, AND ELECTRICAL TESTING. THE ELECTRODE WAS FOUND TO BE SOUND, BOTH INSULATION AND CONTINUITY APPEARED TO BE GOOD. THE ROOT CAUSE WAS DETERMINED TO BE AN ISSUE WITH THE CONNECTING CABLE. A MANUFACTURING DEFECT WAS NOT IDENTIFIED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8080331 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates