Skip to content

ENDOCAM 85525942Recalls & reported problems

Richard Wolf GmbH · LOGIC 4K CAMERA HEAD C-Mount, 3 chip, 2 programmable camera-head buttons, Sensor size: 1/3" · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207063163), received between Mar 19, 2022 and Mar 18, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image2
  • Erratic or Intermittent Display1
  • Excessive Heating1
  • No Visual Prompts/Feedback1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • Burning Sensation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026No Visual Prompts/Feedback

    IT WAS REPORTED THAT DURING A HYSTEROSCOPY PROCEDURE, THE INSTRUMENT, ENDOCAM LOGIC 4K CAMERA HEAD, WAS NOT BEING DETECTED BY THE 4K MEDICAL/VIDEO MONITOR AND FAILED TO DISPLAY A LIVE IMAGE. A 30 MINUTE DELAY OCCURRED WHILE THE STAFF TRIED TO TROUBLESHOOT BUT NONE HAVE RESOLVED THE ISSUE. ANOTHER ENDOCAM CAMERA HEAD WAS ATTAINED AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626844 ↗
  • InjuryDec 22, 2023No Display/Image; No Display/Image

    A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH (RWGMBH) OF AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD, PART ID: 85525942, S/N # (B)(6). ACCORDING TO THE RECEIVED INFORMATION, WHEN SETTING UP THE EQUIPMENT FOR THE ENDOSCOPIC SPINE SURGERY THE STAFF NOTICED THAT THE CAMERA HEAD WAS NOT VISIBLE AND REGISTERED BY THE VIDEO PROCESSOR. MTA AND SERVICE TECHNICIAN FROM RICHARD WOLF DID A SEARCH FOR THE ERROR BUT DID NOT MANAGE TO GET THE CAMERA HEAD TO WORK AT ALL. AS THE PATIENT WAS ALREADY READY FOR THE SURGERY, THE SURGEON TOOK THE DECISION TO SWITCH TO OPEN SURGERY INSTEAD OF PLANNED ENDOSCOPIC SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 18386255 ↗
  • MalfunctionJun 22, 2023Overheating of Device; Overheating of Device

    A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD C-MOUNT, PART ID: 85525942, SERIAL #: (B)(6). ACCORDING TO THE RECEIVED INFORMATION, IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. ALSO, IT WAS REPORTED THAT THIS EXTENDED THE OPERATING ROOM TIME BY 2 HOURS.

    Patient problems: Burning Sensation; Insufficient Information

    MAUDE report 17185875 ↗
  • MalfunctionAug 29, 2022Excessive Heating; Excessive Heating

    RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). RICHARD WOLF GMBH (MANUFACTURER) WAS INFORMED ABOUT THE FOLLOWING: DESCRIPTION OF INCIDENT/CLAIM: IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. AS A RESULT, THE USER HAD TO STOP THE PROCEDURE AND CONTINUE IT THE NEXT DAY. NO INJURIES TO THE PATIENT, USER OR THIRD PARTIES WERE INDICATED. RWGMBH AWARENESS DATE, 01-AUG-2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15314679 ↗
  • MalfunctionMar 19, 2022Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    IT WAS REPORTED TO RICHARD WOLF (B)(4) THAT "THE USER HAS REPORTED THAT THE IMAGE FLICKERED AND FAILED COMPLETELY DURING THE PROCEDURE WITH THE USE OF THE ERBE DEVICE. THERE WAS A DELAY OF 30 MINUTES, THE PROCEDURE COULD BE COMPLETED WITH ANOTHER CAMERA CABLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13827566 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.