ENDOCAM 85525942Recalls & reported problems
Richard Wolf GmbH · LOGIC 4K CAMERA HEAD C-Mount, 3 chip, 2 programmable camera-head buttons, Sensor size: 1/3" · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207063163), received between Mar 19, 2022 and Mar 18, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image2
- Erratic or Intermittent Display1
- Excessive Heating1
- No Visual Prompts/Feedback1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Burning Sensation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2026No Visual Prompts/Feedback
IT WAS REPORTED THAT DURING A HYSTEROSCOPY PROCEDURE, THE INSTRUMENT, ENDOCAM LOGIC 4K CAMERA HEAD, WAS NOT BEING DETECTED BY THE 4K MEDICAL/VIDEO MONITOR AND FAILED TO DISPLAY A LIVE IMAGE. A 30 MINUTE DELAY OCCURRED WHILE THE STAFF TRIED TO TROUBLESHOOT BUT NONE HAVE RESOLVED THE ISSUE. ANOTHER ENDOCAM CAMERA HEAD WAS ATTAINED AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626844 ↗InjuryDec 22, 2023No Display/Image; No Display/Image
A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH (RWGMBH) OF AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD, PART ID: 85525942, S/N # (B)(6). ACCORDING TO THE RECEIVED INFORMATION, WHEN SETTING UP THE EQUIPMENT FOR THE ENDOSCOPIC SPINE SURGERY THE STAFF NOTICED THAT THE CAMERA HEAD WAS NOT VISIBLE AND REGISTERED BY THE VIDEO PROCESSOR. MTA AND SERVICE TECHNICIAN FROM RICHARD WOLF DID A SEARCH FOR THE ERROR BUT DID NOT MANAGE TO GET THE CAMERA HEAD TO WORK AT ALL. AS THE PATIENT WAS ALREADY READY FOR THE SURGERY, THE SURGEON TOOK THE DECISION TO SWITCH TO OPEN SURGERY INSTEAD OF PLANNED ENDOSCOPIC SURGERY.
Patient problems: Insufficient Information
MAUDE report 18386255 ↗MalfunctionJun 22, 2023Overheating of Device; Overheating of Device
A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD C-MOUNT, PART ID: 85525942, SERIAL #: (B)(6). ACCORDING TO THE RECEIVED INFORMATION, IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. ALSO, IT WAS REPORTED THAT THIS EXTENDED THE OPERATING ROOM TIME BY 2 HOURS.
Patient problems: Burning Sensation; Insufficient Information
MAUDE report 17185875 ↗MalfunctionAug 29, 2022Excessive Heating; Excessive Heating
RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). RICHARD WOLF GMBH (MANUFACTURER) WAS INFORMED ABOUT THE FOLLOWING: DESCRIPTION OF INCIDENT/CLAIM: IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. AS A RESULT, THE USER HAD TO STOP THE PROCEDURE AND CONTINUE IT THE NEXT DAY. NO INJURIES TO THE PATIENT, USER OR THIRD PARTIES WERE INDICATED. RWGMBH AWARENESS DATE, 01-AUG-2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15314679 ↗MalfunctionMar 19, 2022Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
IT WAS REPORTED TO RICHARD WOLF (B)(4) THAT "THE USER HAS REPORTED THAT THE IMAGE FLICKERED AND FAILED COMPLETELY DURING THE PROCEDURE WITH THE USE OF THE ERBE DEVICE. THERE WAS A DELAY OF 30 MINUTES, THE PROCEDURE COULD BE COMPLETED WITH ANOTHER CAMERA CABLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13827566 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.