Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207063163), received between Mar 19, 2022 and Mar 18, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image2
- Erratic or Intermittent Display1
- Excessive Heating1
- No Visual Prompts/Feedback1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Burning Sensation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2026No Visual Prompts/Feedback
IT WAS REPORTED THAT DURING A HYSTEROSCOPY PROCEDURE, THE INSTRUMENT, ENDOCAM LOGIC 4K CAMERA HEAD, WAS NOT BEING DETECTED BY THE 4K MEDICAL/VIDEO MONITOR AND FAILED TO DISPLAY A LIVE IMAGE. A 30 MINUTE DELAY OCCURRED WHILE THE STAFF TRIED TO TROUBLESHOOT BUT NONE HAVE RESOLVED THE ISSUE. ANOTHER ENDOCAM CAMERA HEAD WAS ATTAINED AND WAS ABLE TO COMPLETE THE PROCEDURE. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626844 ↗InjuryDec 22, 2023No Display/Image; No Display/Image
A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH (RWGMBH) OF AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD, PART ID: 85525942, S/N # (B)(6). ACCORDING TO THE RECEIVED INFORMATION, WHEN SETTING UP THE EQUIPMENT FOR THE ENDOSCOPIC SPINE SURGERY THE STAFF NOTICED THAT THE CAMERA HEAD WAS NOT VISIBLE AND REGISTERED BY THE VIDEO PROCESSOR. MTA AND SERVICE TECHNICIAN FROM RICHARD WOLF DID A SEARCH FOR THE ERROR BUT DID NOT MANAGE TO GET THE CAMERA HEAD TO WORK AT ALL. AS THE PATIENT WAS ALREADY READY FOR THE SURGERY, THE SURGEON TOOK THE DECISION TO SWITCH TO OPEN SURGERY INSTEAD OF PLANNED ENDOSCOPIC SURGERY.
Patient problems: Insufficient Information
MAUDE report 18386255 ↗MalfunctionJun 22, 2023Overheating of Device; Overheating of Device
A DISTRIBUTOR HAS INFORMED RICHARD WOLF GMBH AN ISSUE REGARDING A LOGIC 4K CAMERA HEAD C-MOUNT, PART ID: 85525942, SERIAL #: (B)(6). ACCORDING TO THE RECEIVED INFORMATION, IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. ALSO, IT WAS REPORTED THAT THIS EXTENDED THE OPERATING ROOM TIME BY 2 HOURS.
Patient problems: Burning Sensation; Insufficient Information
MAUDE report 17185875 ↗MalfunctionAug 29, 2022Excessive Heating; Excessive Heating
RICHARD WOLF GMBH COMPLAINT REFERENCE (B)(4). RICHARD WOLF GMBH (MANUFACTURER) WAS INFORMED ABOUT THE FOLLOWING: DESCRIPTION OF INCIDENT/CLAIM: IT WAS COMPLAINED THAT THE CAMERA HEAD HEATED UP SO MUCH THAT THE USER COULD NO LONGER HOLD IT IN HIS HAND. AS A RESULT, THE USER HAD TO STOP THE PROCEDURE AND CONTINUE IT THE NEXT DAY. NO INJURIES TO THE PATIENT, USER OR THIRD PARTIES WERE INDICATED. RWGMBH AWARENESS DATE, 01-AUG-2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15314679 ↗MalfunctionMar 19, 2022Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image
IT WAS REPORTED TO RICHARD WOLF (B)(4) THAT "THE USER HAS REPORTED THAT THE IMAGE FLICKERED AND FAILED COMPLETELY DURING THE PROCEDURE WITH THE USE OF THE ERBE DEVICE. THERE WAS A DELAY OF 30 MINUTES, THE PROCEDURE COULD BE COMPLETED WITH ANOTHER CAMERA CABLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13827566 ↗
FDA clearance and classification
Cleared through 510(k) K180583 on Jun 28, 2018 (decision: Substantially Equivalent). Submitted by Richard Wolf Medical Instruments Corporation as “Logic 4k Camera Controller, Logic 4k Camera Head”. The FDA review took 115 days. FDA record ↗
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ENDOCAM 85525942
Has the ENDOCAM 85525942 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04055207063163), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ENDOCAM 85525942?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Mar 19, 2022 and Mar 18, 2026. In those reports, the most common specific device problems are “No Display/Image” (2), “Erratic or Intermittent Display” (1) and “Excessive Heating” (1); the most common patient problem is “Burning Sensation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ENDOCAM 85525942?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ENDOCAM 85525942 FDA approved?
It was cleared through 510(k) K180583 on Jun 28, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the ENDOCAM 85525942, and is it still sold?
The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand ENDOCAM, model 85525942). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,392 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page