ENDOCAM 5525301Recalls & reported problems
Richard Wolf GmbH · LOGIC 4K CAMERA CONTROLLER touch screen, color display, compatible with core nova, Input: 1x 3G-SDI (PiP), Output: 2x HDMI 4K, 2x HDMI HD, 2x 3G-SDI, Capturing video/images on USB, Resolution: 3840x2160, 1920x1200, U: 100-240VAC, 50/60Hz, Dim (wxhxd) 300x120x416mm · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04055207063286), received between Jun 23, 2022 and Jul 7, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Erratic or Intermittent Display” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Erratic or Intermittent Display2
- Visual Prompts will not Clear1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2023Visual Prompts will not Clear; Visual Prompts will not Clear
A USER FACILITY HAS INFORMED RICHARD WOLF GMBH AN ISSUE REGARDING A LOGIC 4K CAMERA CONTROLLER, PART ID: 5525301, SERIAL # (B)(6). ACCORDING TO THE RECEIVED INFORMATION, AT THE BEGINNING OF A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) PROCEDURE, THE VIDEO PROCESSOR WAS SWITCHED ON. THE OPTICS WERE ALREADY IN THE PATIENT'S URETHRA. AFTER COMPLETION OF THE VIDEO PROCESSOR'S SELF-TEST, A LARGE YELLOW BANNER WAS DISPLAYED ON THE MONITOR: "THE DEVICE MAY HAVE LIMITED FUNCTIONALITY. NOTIFY SERVICE! ERROR..." ALTHOUGH THE CAMERA DISPLAYED AN IMAGE ON THE MONITOR, IT WAS ALMOST COMPLETELY OBSCURED BY THE ERROR MESSAGE. CONFIRMATION / ACKNOWLEDGEMENT WAS NOT POSSIBLE, SO THE APPLICATION HAD TO BE INTERRUPTED AND A REPLACEMENT DEVICE HAD TO BE INTEGRATED.
Patient problems: Insufficient Information
MAUDE report 17280831 ↗MalfunctionJun 23, 2022Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED TO RWMIC (IMPORTER) THAT "THE SCREEN GOES IN AND OUT DURING CASE. NOT SURE IF IT'S THE CCU/MONITOR/STAND." THROUGH EMAIL, THE FOLLOWING INFORMATION WAS PROVIDED; ((B)(6) 2022): PAPERWORK FOR A LOANER DEVICE WAS SENT. ((B)(6) 2022): THE PROCEDURE WAS DELAYED 45 MINUTES. ANOTHER COMPANY'S VIDEO EQUIPMENT WAS BROUGHT IN DURING THE DELAY. THE PATIENT WAS GIVEN MORE ANESTHESIA DURING THE DELAY. THE PATIENT OUTCOME WAS STENT PLACEMENT. ((B)(6) 2022): CASE SCHEDULE AS AN URETEROSCOPY STENT PLACEMENT. THE PATIENT WAS GIVEN ADDITIONAL MEDICATION [ANESTHESIA] WHILE THE ISSUE WAS BEING RESOLVED, WHICH ALLOWS FOR A POTENTIAL RISK. WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTED ISSUE OCCURRED? YES WAS. THERE INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO DID. THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? INITIALLY REPORTED AS NO. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. MDR AWARENESS DATE: 05/31/2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14787102 ↗MalfunctionJun 23, 2022Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED TO RWMIC THAT "THE SCREEN GOES IN AND OUT DURING CASE. NOT SURE IF IT'S THE CCU/MONITOR/STAND." THROUGH EMAIL, THE FOLLOWING INFORMATION WAS PROVIDED; (05/25/2022): PAPERWORK FOR A LOANER DEVICE WAS SENT. (05/27/2022): THE PROCEDURE WAS DELAYED 45 MINUTES. ANOTHER COMPANY'S VIDEO EQUIPMENT WAS BROUGHT IN DURING THE DELAY. THE PATIENT WAS GIVEN MORE ANESTHESIA DURING THE DELAY. THE PATIENT OUTCOME WAS A STENT PLACEMENT. (05/31/2022): THE PATIENT WAS GIVEN ADDITIONAL MEDICATION [ANESTHESIA] WHILE THE ISSUE WAS BEING RESOLVED, WHICH ALLOWS FOR A POTENTIAL RISK. CASE SCHEDULE AS AN URETEROSCOPY STENT PLACEMENT. WILL THE DEVICE BE RETURNED? YES. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTED ISSUE OCCURRED? YES. WAS THERE INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? YES. DID THE DELAY PUT THE PATIENT AT RISK? INITIALLY REPORTED AS NO. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES. RWMIC MDR AWARENESS DATE: (B)(6) 2022.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14787091 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.