SILVERGLIDE 6720220012Recalls & reported problems

Stryker Corporation · Non-Stick Bipolar Electrosurgical Precision Forceps Bayonet · Class 2

Quick summary

This Stryker Silverglide bipolar electrosurgical forceps is used for cutting and coagulating tissue during surgery. The 34 reports mostly describe the instrument wearing apart — tips or the guide stop breaking off, and the silver plating flaking away, which the reports say could leave pieces of material at the surgical site. No deaths and no injuries were reported. There are no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 04546540532299), received between Aug 16, 2018 and Jul 7, 2025 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component19
  • Flaked13
  • Fracture1
  • Material Fragmentation1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Patient Involvement6

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22419869 ↗
  • MalfunctionJun 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22359637 ↗
  • MalfunctionJun 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22359041 ↗
  • MalfunctionJun 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22358725 ↗
  • MalfunctionJun 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22153231 ↗
  • MalfunctionMay 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22106714 ↗
  • MalfunctionMay 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THE GUIDE STOP WAS FRACTURED OFF AND MISSING FROM THE DEVICE DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22098369 ↗
  • MalfunctionMay 22, 2025Flaked; Flaked

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THE SILVER PLATING WAS FLAKED OFF THE TIPS OF THE DEVICE, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069804 ↗
  • MalfunctionMay 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE GUIDE STOP WAS OBSERVED TO BE BROKEN OFF THE DEVICE DURING PROCESSING IN SPD. THERE WAS NO PATIENT INVOLVEMENT, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22001908 ↗
  • MalfunctionMar 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THE GUIDE STOP WAS FRACTURED OFF AND MISSING FROM THE DEVICE DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21690695 ↗
  • MalfunctionMar 5, 2025Flaked; Flaked

    IT WAS REPORTED DURING TESTING AT THE USER FACILITY, PIECES OF INSULATION WERE CHIPPED OFF THE TIPS OF THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21524916 ↗
  • MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED ONE OF THE TIPS OF THE DEVICE WAS OBSERVED TO BE FRACTURED OFF AND MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21211376 ↗
  • MalfunctionAug 9, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING EVALUATION AT THE MANUFACTURER FACILITY, THE DEVICE TIPS WERE FOUND TO BE CHIPPED AND SILVER PLATING MATERIAL WAS MISSING, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19955400 ↗
  • MalfunctionAug 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE GUIDE STOP WAS FOUND TO BE DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19953684 ↗
  • MalfunctionMar 4, 2024Flaked; Flaked

    IT WAS REPORTED THAT DURING CLEANING AT THE USER FACILITY, PIECES OF INSULATION WERE FOUND TO BE MISSING FROM THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18826045 ↗
  • MalfunctionMay 26, 2023Flaked; Flaked

    IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE INSULATION WAS FOUND TO BE COMING OFF THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011159 ↗
  • MalfunctionMar 2, 2023Flaked; Flaked

    THE USER FACILITY REPORTED THE DEVICE INSULATION WAS COMING OFF, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16473084 ↗
  • MalfunctionDec 22, 2022Flaked; Flaked

    THE USER FACILITY REPORTED THE INSULATION WAS COMING OF THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16036505 ↗
  • MalfunctionNov 16, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE INSULATION WAS FOUND TO BE COMING OFF, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15809609 ↗
  • MalfunctionNov 14, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE INSULATION WAS FOUND TO BE COMING OFF, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15790719 ↗
  • MalfunctionNov 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE GUIDE STOP WAS FOUND TO BE DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING TESTING BY A MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15779580 ↗
  • MalfunctionJul 13, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE TIP WAS FOUND TO BE FRACTURED OFF AND MISSING FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, IN THE STERILE PROCESS DEPARTMENT OF THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15009263 ↗
  • MalfunctionApr 22, 2022Flaked; Flaked

    THE USER FACILITY REPORTED THE DEVICE INSULATION WAS COMING OFF AT THE TIPS, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14190304 ↗
  • MalfunctionMar 28, 2022Flaked; Flaked

    IT WAS REPORTED THAT DURING TESTING IN THE STERILE PROCESSING DEPARTMENT OF THE USER FACILITY, INSULATION WAS PEELING OFF NEAR THE TIPS OF THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13929525 ↗
  • MalfunctionNov 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE GUIDE STOP WAS FOUND TO BE DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12857760 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →