Skip to content

SILVERGLIDE 6720240015Recalls & reported problems

Stryker Corporation · Non-Stick Bipolar Electrosurgical Precision Forceps Bayonet · Class 2

All models of SILVERGLIDE (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540532343), received between Mar 15, 2019 and Feb 16, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Flaked” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Flaked3
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2021Flaked; Flaked

    IT WAS REPORTED THAT DURING TESTING AT THE USER FACILITY, INSULATION WAS FOUND TO BE MISSING FROM THE DEVICE, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11334978 ↗
  • MalfunctionNov 18, 2020Flaked; Flaked

    IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, THE COATING OF THE DEVICE WAS COMING OFF, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10862769 ↗
  • MalfunctionNov 18, 2020Flaked; Flaked

    IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, PIECES OF INSULATION WERE FOUND TO BE MISSING FROM THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10862660 ↗
  • MalfunctionMar 15, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING INSPECTION AT THE USER FACILITY, INSULATION WAS FOUND TO BE COMING OFF THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8425912 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices