SilverGlide 6730220012Recalls & reported problems
Stryker Corporation · ProSeries Non-stick Bipolar Electrosurgical Forceps Bayonet Tip · Class 2
All models of SILVERGLIDE (60)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540695956), received between Dec 22, 2020 and Dec 15, 2022 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Flaked3
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2022Flaked; Flaked
IT WAS REPORTED THAT DURING EVALUATION AT THE MANUFACTURER FACILITY, THE DEVICE TIPS WERE FOUND TO BE CHIPPED AND SILVER PLATING MATERIAL WAS MISSING, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15988885 ↗MalfunctionAug 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING TESTING BY THE MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY, A PIECE OF THE DEVICE INSULATION WAS MISSING, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15190187 ↗MalfunctionDec 7, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
DURING TESTING AT THE USER FACILITY BY THE MANUFACTURER SALES REPRESENTATIVE, IT WAS REPORTED A PIECE OF INSULATION WAS CHIPPED OFF THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12956824 ↗MalfunctionOct 6, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE TIP WAS FOUND TO BE FRACTURED OFF AND MISSING FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING TESTING BY THE MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12586533 ↗MalfunctionOct 6, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE TIP WAS FOUND TO BE FRACTURED OFF AND MISSING FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING TESTING BY THE MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12586429 ↗MalfunctionDec 22, 2020Flaked; Flaked
IT WAS REPORTED THAT DURING EVALUATION BY THE MANUFACTURER SALES REPRESENTATIVE IN THE STERILE PROCESSING DEPARTMENT AT THE USER FACILITY, PIECES OF INSULATION WERE FOUND TO BE MISSING FROM THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11058078 ↗MalfunctionDec 22, 2020Flaked; Flaked
IT WAS REPORTED THAT DURING EVALUATION BY A MANUFACTURER SALES REPRESENTATIVE IN THE STERILE PROCESSING DEPARTMENT AT THE USER FACILITY, PIECES OF INSULATION WERE FOUND TO BE MISSING FROM THE DEVICE, POSING THE RISK OF A MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11057883 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page