Skip to content

SILVERGLIDE 6720220020Recalls & reported problems

Stryker Corporation · Non-Stick Bipolar Electrosurgical Precision Forceps Bayonet · Class 2

All models of SILVERGLIDE (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540714923), received between May 13, 2019 and Feb 12, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Flaked1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2025Flaked; Flaked

    IT WAS REPORTED DURING TESTING AT THE USER FACILITY MATERIAL WAS OBSERVED TO BE PEELED OFF THE TIPS OF THE DEVICE, POSING THE RISK OF MATERIAL BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21365900 ↗
  • MalfunctionMay 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE GUIDE STOP WAS FOUND TO BE DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19430245 ↗
  • MalfunctionJul 27, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING TESTING AFTER FIRST USE AT THE USER FACILITY, IT WAS REPORTED THE DEVICE GUIDE STOP SPLIT AND DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17406201 ↗
  • MalfunctionMay 13, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE GUIDE STOP BUSHING WAS FOUND TO BE DISASSEMBLED FROM THE DEVICE, POSING THE RISK OF A SMALL COMPONENT BEING LOST IN A SURGICAL SITE, DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8604655 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page