A CORD MAJ-860Recalls & reported problems

Olympus Medical Systems Corp. · "MAJ-860 A-CORD FOR VALLEY ERBE BOVIE" · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170062872), received between Jul 19, 2021 and Jul 19, 2021 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Electric Shock2
  • Numbness2
  • Inflammation1
  • Swelling/ Edema1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT

    Patient problems: Electric Shock; Numbness; Swelling/ Edema

    MAUDE report 12188021 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTS WHILE PREPARING FOR AN UNSPECIFIED PROCEDURE USING A WM-DP2 WORKSTATION AND A MAJ-860 ACTIVE CORD (ALONG WITH A VALLEYLAB GENERATOR AND BOVIE), THE ENDOSCOPY TECHNICIAN EXPERIENCED AN ELECTRICAL SHOCK WHILE CONNECTING AN ELECTROSURGERY ACCESSORY (BOSTON SCIENTIFIC SINGLE USE POLYPECTOMY SNARE). IT IS REPORTED THE TECHNICIAN'S GLOVES WERE WET AND THE GENERATOR WAS POWERED ON AT THE TIME HE ATTEMPTED TO CONNECT THE SNARE. THE TECHNICIAN DECLINED AN EMERGENCY ROOM VISIT FOR EVALUATION AT THE TIME AND CONTINUED TO WORK THE REMAINDER OF THE DAY. THE TECHNICIAN'S CURRENT CONDITION IS DESCRIBED AS UNABLE TO WORK DUE TO "HIS ARM AND HAND ARE STILL SWOLLEN AND NUMB." THE DEVICES WERE USED FOR THE REMAINED OF THE DAY WITH NO FURTHER ISSUES REPORTED. BIOMED CHECKED UNITS AND FOUND NO ISSUES. THE MAJ-860 WAS INSPECTED WITH NO VISIBLE DAMAGE NOTED. BIOMED HAS REMOVED THE NON-OLYMPUS GENERATOR FROM USE AND REPLACED WITH ANOTHER UNIT ON A SEPARATE CART/ ROLL STAND. THE OLYMPUS CORD (MAJ-860) AND BOVIE WERE CHECKED OUT BY BIOMED, AND NO MALFUNCTIONS WERE IDENTIFIED. THE OLYMPUS MAJ-860 HAS BEEN REPLACED WITH A BOSTON SCIENTIFIC ACTIVE CORD. NO FURTHER ISSUES REPORTED. CASE WIT

    Patient problems: Electric Shock; Inflammation; Numbness

    MAUDE report 12187951 ↗