EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F (UDI 04953170066344)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS DUODENOVIDEOSCOPE (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170066344), received between Dec 7, 2018 and May 3, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Microbial Contamination of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Microbial Contamination of Device2
- Crack1
- Device Reprocessing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 3, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD BEEN REPROCESSED WITH AN ENDOSCOPE REPROCESSOR IN WHICH THE WATER FILTER REPLACEMENT DEADLINE HAD EXPIRED BY TWO YEARS. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19237934 ↗MalfunctionJan 10, 2023Crack; Crack
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS DUODENOVIDEOSCOPE PLASTIC DISTAL END COVER WAS CRACKED. ALTHOUGH REQUESTED, THE CUSTOMER WAS UNABLE TO PROVIDE FURTHER DETAILS. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16128717 ↗MalfunctionJul 11, 2019Microbial Contamination of Device; Microbial Contamination of Device
THE MANUFACTURE WAS INFORMED THAT DURING A POST MARKET STUDY THE SCOPE CULTURED POSITIVE FOR KLEBSIELLA PNEUMONIAE AFTER REPROCESSING. THERE WAS NO ASSOCIATED PATIENT INFECTION REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8781821 ↗MalfunctionDec 7, 2018Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS WAS INFORMED THAT DURING A POST MARKET SURVEILLANCE STUDY THE SCOPE CULTURED POSITIVE FOR A HIGH CONCERN MICROORGANISM, ESCHERICHIA COLI / SHIGELLA DYSENTERIAE, AFTER REPROCESSING. THERE WERE NO ASSOCIATED PATIENT INFECTIONS REPORTED. THE OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) REPORTED AN IN-SERVICE ON REPROCESSING WAS LAST PERFORMED BY AT THE USER FACILITY ON NOVEMBER 2, 2018 AND ON NOVEMBER 5, 2018. ALL REPROCESSING PERSONNEL WERE PROPERLY TRAINED. HOWEVER, THE TECHNICIAN WHO REPROCESSED THIS SCOPE HAS NOT BEEN IN-SERVICED BY OLYMPUS AND WAS TRAINED BY THEIR STAFF. THE STAFF IS FOLLOWING THE INSTRUCTION MANUAL'S REPROCESSING RECOMMENDATIONS. IN ADDITION, PRE-CLEANING IS BEING PERFORMED IMMEDIATELY AFTER A PROCEDURE. THE ENDOSCOPE IS BEING LEAK TESTED PRIOR TO MANUAL CLEANING. THE ENDOSCOPE CHANNEL IS BRUSHED DURING MANUAL CLEANING. THE MINIMUM EFFECTIVE CONCENTRATION IS CHECKED BEFORE EVERY SCOPE. THE NON-OLYMPUS AER HAS NO REPORTED ISSUES AND IT IS UNKNOWN WHEN THE LAST MAINTENANCE WAS PERFORMED. THE SCOPE IS STORED IN A STORAGE CABINET, HANGING VERTICALLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8143228 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page