EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE 2T160 (UDI 04953170066610)Recalls & reported problems

Olympus Medical Systems Corp. · GIF-2T160 THERAPEUTIC VIDEO GASTROSCOPE · Class 2

All models of EVIS EXERA GASTROINTESTINAL VIDEOSCOPE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus GIF-2T160 therapeutic video gastroscope is a flexible scope used to examine and treat the stomach and upper digestive tract. Most of the 32 reports describe cleaning and reprocessing problems — residue, black or red material, mold and bacteria found inside the air/water nozzle or channels, along with a few images that failed to appear. None report a death or an injury. There are no recalls name this device, and reports in the last 12 months, at 16, were about twice the 8 in the prior year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 04953170066610), received between Jun 13, 2018 and Jun 8, 2026 — classified as 32 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 7 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately11
  • Microbial Contamination of Device4
  • Device Reprocessing Problem3
  • Poor Quality Image3
  • Residue After Decontamination3
  • Erratic or Intermittent Display2
  • Melted2
  • No Display/Image2

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Consequences Or Impact To Patient2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD FOREIGN MATERIAL IN THE AIR/WATER NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25434067 ↗
  • MalfunctionMay 13, 2026Failure to Clean Adequately

    THE GASTROINTESTINAL VIDEOSCOPE WAS RETURNED TO OLYMPUS WITH A REPORTED COMPLAINT OF FAILED LEAK TEST THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER, THERE WERE FOREIGN MATERIALS COMING OUT OF THE DISTAL END OF THE GASTROINTESTINAL VIDEOSCOPE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162465 ↗
  • MalfunctionMay 5, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE NOZZLE OF THE GASTROINTESTINAL VIDEOSCOPE HAD BLACK FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084546 ↗
  • MalfunctionApr 17, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE NOZZLE OF THE GASTROINTESTINAL VIDEOSCOPE HAD RESIDUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24910139 ↗
  • MalfunctionApr 9, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE NOZZLE OF THE GASTROINTESTINAL VIDEOSCOPE HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24827296 ↗
  • MalfunctionMar 12, 2026Failure to Clean Adequately

    IT WAS OBSERVED DURING DEVICE EVALUATION, THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A RED SUBSTANCE ON THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24578828 ↗
  • MalfunctionFeb 6, 2026Microbial Contamination of Device

    IT WAS REPORTED THAT DURING REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR >100 COLONY FORMING UNITS (CFU) OF CITROBACTER FREUNDII, GRAM-NEGATIVE BACILLI OF THE ENTEROBACTERIACEAE FAMILY, AND MOLDS. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24281608 ↗
  • MalfunctionJan 22, 2026No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24151446 ↗
  • MalfunctionDec 29, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23918837 ↗
  • MalfunctionDec 23, 2025Residue After Decontamination; Residue After Decontamination

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE RESIDUE IN THE AIR/WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23882753 ↗
  • MalfunctionDec 12, 2025Residue After Decontamination; Residue After Decontamination

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED CHEMICAL RESIDUE IN THE AUXILIARY CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 23791144 ↗
  • MalfunctionDec 10, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE OBJECT STUCK INSIDE THE SCOPE, THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23769610 ↗
  • MalfunctionDec 9, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROVIDEOSCOPE IMAGE-EVIS EXHIBITED INTERMIT/FLICKER IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23753217 ↗
  • MalfunctionDec 8, 2025Microbial Contamination of Device; Microbial Contamination of Device

    NO ADDITIONAL INFORMATION FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23752574 ↗
  • MalfunctionDec 2, 2025Erratic or Intermittent Display; Poor Quality Image; Erratic or Intermittent Display; Poor Quality Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BLURRY AND FLICKERING IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23697583 ↗
  • MalfunctionNov 19, 2025Melted; Melted

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE'S CAMERA COVER WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23596603 ↗
  • MalfunctionOct 28, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A LARGE AMOUNT OF GLUE ON THE PLASTIC DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23405598 ↗
  • MalfunctionSep 9, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE WAS OBSERVED TO HAVE FOREIGN OBJECT STUCK INSIDE THE BIOPSY CHANNEL MOUNT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23010561 ↗
  • MalfunctionAug 1, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD SUTURE STUCK IN SCOPE. THE EVENT HAD OCCURRED DURING THE THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22683864 ↗
  • MalfunctionJul 10, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THE SUBJECT DEVICE HAD SOMETHING STUCK IN THE CHANNEL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22473292 ↗
  • MalfunctionJun 26, 2025Melted; Failure to Clean Adequately; Melted; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FOREIGN MATERIAL IN THE AIR-WATER CHANNEL AND THE JET TUBE, ALSO THE CAMERA COVER WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22341361 ↗
  • MalfunctionJun 4, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE UNIVERSAL CORD WAS STICKY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22140423 ↗
  • MalfunctionMay 27, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE WAS TRANSPORTED IN A BUCKET THAT CONTAINED A SIGNIFICANT AMOUNTS OF WATER WITH NO CAP IN PLACE. THIS OCCURRED WHILE THE SCOPE WAS BEING CLEANED AT THE BEDSIDE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22087456 ↗
  • MalfunctionMay 12, 2025Residue After Decontamination; Residue After Decontamination

    IT WAS OBSERVED THAT DURING DEVICE REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED BLACK RESIDUE LEAKING FROM THE DISTAL END OF SCOPE AFTER HIGH-LEVEL DISINFECTION (HDL) PROCESSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22003254 ↗
  • MalfunctionJul 24, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED, THE FLEX VIDEO SCOPE EXHIBITED CLOGGED NOZZLE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822246 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →