Distal cover MAJ-311Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04953170071393), received between Oct 12, 2021 and Nov 15, 2024 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Use of Device Problem2
- Break1
- Device Fell1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Laceration(s)1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED, THE DISTAL COVER INTERFERED WITH THE IMAGE DUE TO DISTAL CAP VISIBILITY, THE CAP COULD NOT BE SEPARATED FROM THE SCOPE AND WHEN REMOVED USING LIQUID A SCORCHED AREA WAS NOTED INSIDE THE CAP. THE ISSUE OCCURRED DURING A THERAPEUTIC ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY) PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THERE WAS NO DELAY IN PROCEDURE. THE PROCEDURE WAS COMPLETED NORMALLY WITH ANOTHER OF THE SAME MODEL OF DEVICE. THE PRODUCT WAS INSPECTED BEFORE THE PROCEDURE AND THERE WAS NO ABNORMALITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20695564 ↗InjuryAug 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT AT THE END OF THERAPEUTIC, ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) STENT EXCHANGE PROCEDURE, THE DISTAL COVER FELL OUT INTO THE PATIENT¿S ESOPHAGUS WHILE THE DUODENOVIDEOSCOPE WAS BEING PULLED OUT OF THE BODY. THE DISTAL COVER WAS RETRIEVED WITH THE AID OF GRASPING FORCEPS. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO MUCOSAL DAMAGE, AND NO HARM TO THE PATIENT. THE COVER WAS DISPOSED OF BY THE FACILITY. THE DEVICE WAS INSPECTED PRIOR TO USE BUT THE INSPECTION MAY HAVE BEEN INCOMPLETE. AN OLYMPUS REPRESENTATIVE SUBSEQUENTLY EXPLAINED THE CORRECT INSTALLATION AND INSPECTION METHOD TO THE STAFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19954493 ↗MalfunctionJun 19, 2024Detachment of Device or Device Component; Device Fell; Detachment of Device or Device Component; Device Fell
IT WAS REPORTED, THAT THE DISTAL COVER FELL OFF AFTER INSTALLATION. THIS ISSUE OCCURRED AFTER A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19570759 ↗InjuryMay 2, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING REPROCESSING AFTER USING A DUODENOVIDEOSCOPE, DAMAGE TO THE DISTAL COVER WAS CONFIRMED. SINCE THE POSSIBILITY OF DAMAGE OCCURRING INSIDE THE BODY DURING THE UNSPECIFIED SURGERY COULD NOT BE RULED OUT, AN ENDOSCOPIC EXAMINATION OF THE UPPER GASTROINTESTINAL TRACT WAS PERFORMED AGAIN TO CONFIRM THAT NO RUBBER FRAGMENTS WERE PRESENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19229979 ↗MalfunctionNov 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE DISTAL COVER END WAS MISSING. THE EVENT OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18056892 ↗MalfunctionFeb 24, 2022Use of Device Problem; Use of Device Problem
DURING A PROCEDURE, THE DISTAL COVER FELL OFF THE END OF THE SCOPE INTO THE PATIENTS STOMACH. THE COVER WAS RETRIEVED, AND ALTHOUGH THERE WAS A SLIGHT DELAY, THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WAS NO PATIENT HARM AS A RESULT OF THE EVENT. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE JF-260V FLEX VIDEO SCOPE USED IN THE PROCEDURE. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE MAJ-311 DISTAL COVER USED IN THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592404 ↗InjuryOct 12, 2021Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTS DURING AN UNSPECIFIED PROCEDURE USING AN EVIS EXERA DUODENOVIDEOSCOPE WITH A DISPOSABLE DISTAL CAP, THE PATIENT EXPERIENCED A MUCOSAL LACERATION. NO FURTHER DETAILS REGARDING THE PATIENT OR EVENT COULD BE PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE SCOPE USED IN THE CASE. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS THE DISTAL CAP USED IN THE CASE (THIS CASE).
Patient problems: Laceration(s)
MAUDE report 12616254 ↗
FDA clearance and classification
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Distal cover MAJ-311
Has the Distal cover MAJ-311 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04953170071393), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Distal cover MAJ-311?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Oct 12, 2021 and Nov 15, 2024. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (4), “Use of Device Problem” (2) and “Break” (1); the most common patient problem is “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Distal cover MAJ-311?
None of the reports are classified as deaths, 3 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Distal cover MAJ-311, and is it still sold?
The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand Distal cover, model MAJ-311). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,392 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page