Gas/Water Valve MAJ-521Recalls & reported problems
Olympus Medical Systems Corp. · MAJ-521 GAS/WATER VALVE FOR CF-140/160/1 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170071478), received between Sep 4, 2019 and Jun 19, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Contamination” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Contamination1
- Increase in Pressure1
- Pressure Problem1
Patient problems
- Bowel Perforation1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
- Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2022Contamination; Contamination
THE CUSTOMER REPORTED THE AIR SUPPLY WAS INSUFFICIENT AND WATER REMAINED ON THE LENS SURFACE. IT WAS ALSO REPORTED THAT THE RUBBER PART APPEARED TO BE MISSING. THE REPORTED ISSUE OCCURRED WHILE PREPARING THE SUBJECT DEVICE FOR USE DURING A DIAGNOSTIC COLONOSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT. THE DEVICE WAS RETURNED TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THE VALVE WAS CLOGGED WITH FOREIGN MATERIAL. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION OF THE DEVICE (FOREIGN MATERIAL IN VALVE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14741378 ↗InjurySep 4, 2019Pressure Problem; Pressure Problem
THE MANUFACTURER BECAME INFORMED THAT 24 HOURS AFTER A DIAGNOSTIC COLONOSCOPY PROCEDURE IN (B)(6) 2019, A PATIENT EXPERIENCED PAIN AND WAS INSPECTED. THE DOCTOR AT THE USER FACILITY INDICATED THAT THE PATIENT WAS DETERMINED TO HAVE A PERFORATED RECTUM AND THE DOCTOR BELIEVED THE INJURY TO THE PATIENT WAS CAUSED BY OVER PRESSURE FROM THE ENDOSCOPIC CO2 REGULATION UNIT AND GAS/WATER VALVE. THE PATIENT WAS ADMITTED TO A DIFFERENT HOSPITAL/SURGEON AND REQUIRED A EXPLORATORY LAPAROTOMY. THE PATIENT WAS HOSPITALIZED FOR A FEW WEEKS AND DISCHARGED AS THE PATIENT ALSO HAD A PRE-EXISTING CONDITION OF SEVERE ARTERIAL VASCULAR DISEASE IN THE LOWER EXTREMITIES THAT EVENTUALLY REQUIRED AMPUTATION OF ONE HIS LEGS. THE TECHNICIANS AT THE USER FACILITY PERFORMED A FUNCTIONAL INSPECTION PRIOR TO PROCEDURE AND THERE WERE NO ANOMALIES NOTED. THE DOCTOR NOTICED THAT THERE WAS PRESSURE BEING LOST AT THE TIME BUT WASN'T SURE WHAT IT WAS, BUT THE DOCTOR LATER BELIEVED IT TO BE PRESSURE BEING LOST FROM THE PERFORATION TO THE PATIENT. THIS IS FOR REPORT 2 OF 2.
Patient problems: Bowel Perforation
MAUDE report 8961074 ↗InjurySep 4, 2019Increase in Pressure
THE MANUFACTURER BECAME INFORMED THAT 24 HOURS AFTER A DIAGNOSTIC COLONOSCOPY PROCEDURE IN (B)(6) 2019, A PATIENT EXPERIENCED PAIN AND WAS INSPECTED. THE DOCTOR AT THE USER FACILITY INDICATED THAT THE PATIENT WAS DETERMINED TO HAVE A PERFORATED RECTUM AND THE DOCTOR BELIEVED THE INJURY TO THE PATIENT WAS CAUSED BY OVER PRESSURE FROM THE ENDOSCOPIC CO2 REGULATION UNIT AND GAS/WATER VALVE. THE PATIENT WAS ADMITTED TO A DIFFERENT HOSPITAL/SURGEON AND REQUIRED A EXPLORATORY LAPAROTOMY. THE PATIENT WAS HOSPITALIZED FOR A FEW WEEKS AND DISCHARGED AS THE PATIENT ALSO HAD A PRE-EXISTING CONDITION OF SEVERE ARTERIAL VASCULAR DISEASE IN THE LOWER EXTREMITIES THAT EVENTUALLY REQUIRED AMPUTATION OF ONE HIS LEGS. THE TECHNICIANS AT THE USER FACILITY PERFORMED A FUNCTIONAL INSPECTION PRIOR TO PROCEDURE AND THERE WERE NO ANOMALIES NOTED. THE DOCTOR NOTICED THAT THERE WAS PRESSURE BEING LOST AT THE TIME BUT WASN'T SURE WHAT IT WAS, BUT THE DOCTOR LATER BELIEVED IT TO BE PRESSURE BEING LOST FROM THE PERFORATION TO THE PATIENT. THIS IS FOR REPORT 1 OF 2.
Patient problems: Pain; Perforation
MAUDE report 8961023 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page